Electrical Power Standards
Unpacking and Inspection
7
Environment
Operating temperature
5
°
C - 35
°
C
Calibration temperature (tcal) range
16
°
C - 30
°
C
Storage temperature
0
°
C - 50
°
C
Transit temperature
-20
°
C - 60
°
C <100 hours
Warm up time
1 hour
Safe operating max. relative humidity (non-condensing)
<80 % 5
°
C - 31
°
C ramping linearly down to 50 % at 35
°
C
Storage max. relative humidity (non-condensing)
<95 % 0
°
C - 50
°
C
Operating altitude
0 m - 2,000 m
Non-operating altitude
0 m - 12,000 m
Shock
MIL-PRF-28800F class 3
Vibration
MIL-PRF-28800F class 3
Enclosure
MIL-PRF-28800F class 3
Safety
•
Meets CAN/CSA-C22.2 No 61010.1-04, UL Std. No. 61010-1 (2
nd
Edition), ISA-82.02.01
Reference Standard: IEC 61010-1:2001
•
Indoor use only, pollution degree 2; installation category II
•
CE marked and CSA listed
EMC
EN61326: 2006, CISPR 11 class A, FCC rules part 15, sub-part B, class A (Class A equipment is suitable for use in
establishments other than domestic, and those directly connected to a low voltage power supply network which supplies
buildings used for domestic purposes).
Unpacking and Inspection
The Instruments are shipped in a container designed to prevent damage during shipping.
Inspect the Instrument carefully for damage, and immediately report any damage to the
shipper. Instructions for inspection and claims are included in the shipping container.
A packing list is included in the packaging. When you unpack the Instrument, check for
all the standard equipment listed and check the shipping order for any additional items
ordered. Report any shortage to the place of purchase or to the nearest Fluke Service
Center.
Reshipping the Instrument
A transit case can be purchased from Fluke. The Fluke part number is 1887580. This
container is suitable for most handling conditions but provides less shock protection than
the original cardboard packaging. It is recommended that the original container be used
when possible.
Placement and Rack Mounting
This Instrument is designed to operate in controlled electromagnetic environments such
as calibration and measurement laboratories i.e. where R.F. transmitters such as mobile
telephones are not be used in close proximity.
The Instrument is suitable for bench top use, as long as there is sufficient space on either
side to allow adequate ventilation.
The Instrument can be rack mounted using Fluke part number 1887571. Details of the
rack mounting kit and fitting instructions are provided with the kit. Note that the airflow
through the Instrument is from left to right as viewed from the front. If mounted in a rack,
the airflow must be in the same direction.