Introduction and Specifications
Secondary Electrical Specifications
1
1-15
Voltage
Range
Frequency Range
24 Hour AC
Stability
±
1
°
C
Slow Filter Peak-
Peak
±
μ
V
Temperature Coefficient
[1]
Input
Resistance
[2]
10
°
C to 40
°
C
0
°
C to 10
°
C
40
°
C to 50
°
C
ppm /
°
C
220 V
[3]
10 Hz - 20 Hz
8
1.5
3.0
20 Hz - 40 Hz
5
1.5
3.0
40 Hz - 20 kHz
5
1.5
3.0
20 kHz - 50 kHz
5
2.0
3.0
50 k
Ω
50 kHz - 100 kHz
18
5.0
8.0
100 kHz - 300 kHz
36
15.0
15.0
300 kHz - 500 kHz
48
40.0
40.0
700 V
10 Hz - 20 Hz
[4]
8
1.5
4.0
20 Hz - 40 Hz
5
1.5
4.0
40 Hz – 20 kHz
5
1.5
4.0
500 k
Ω
20 kHz – 50 kHz
18
5.0
7.0
50 kHz - 100 kHz
36
15.0
15.0
1000 V
10 Hz - 20 Hz
[4]
8
1.5
4.0
20 Hz - 40 Hz
5
1.5
4.0
40 Hz - 20 kHz
5
1.5
4.0
500 k
Ω
20 kHz - 50 kHz
[5]
18
5.0
7.0
50 kHz - 100 kHz
[5]
36
15.0
15.0
[1]
Add to uncertainty when more than 5
°
C from calibration temperature.
[2]
Input capacitance approximately 100 pF.
[3]
Inputs with a V*Hz product >2.2 E7 are unspecified.
[4]
Typical specification, as determined by sourcing with the Fluke 5205A Precision Power Amplifier.
[5]
Inputs that are >30 kHz and >750 V are typical, as determined by sourcing with the Fluke 5205A Precision Power Amplifier.
DC Secondary Performance
Voltage Range
Temperature Coefficient
[1]
Input
Resistance
[2]
10
°
C to 40
°
C
0
°
C to 10
°
C
40
°
C to 50
°
C
ppm /
°
C
220 mV
1.5
3.0
10 M
Ω
700 mV
1.5
3.0
10 M
Ω
2.2 V
1.5
3.0
10 M
Ω
7 V
1.5
3.0
50 k
Ω
22 V
1.5
3.0
50 k
Ω
70 V
1.5
3.0
50 k
Ω
220 V
1.5
3.0
50 k
Ω
700 V
1.5
4.0
500 k
Ω
1000 V
1.5
4.0
500 k
Ω
[1]
Add to uncertainty when more than 5
°
C from calibration temperature.
[2]
Input capacitance approximately 100 pF.
Note: DC specification valid only when dc input signal is averaged with an equal and opposite dc input signal to eliminate dc offset
errors. The use of Input 1 for dc inputs is not recommended due to the inherent thermal EMFs in a “N” connector. See Operators
Manual for details.
Summary of Contents for 5790B
Page 6: ...5790B Service Manual iv...
Page 26: ...5790B Service Manual 1 20...
Page 104: ...5790B Service Manual 3 66...
Page 109: ...Maintenance Introduction 4 4 5 hvi026 eps Figure 4 3 Air Filter Access...
Page 132: ...5790B Service Manual 5 10...
Page 134: ...5790B Service Manual...
Page 150: ...5790B Service Manual B 6...