Specifications

Operating Fluid Compatibility

The standard system uses Viton O-ring seals and is for use with a wide range of fluids. 

Compatible fluids include, but are not limited to, mineral oils, water, and solvents. For 

compatibility with phosphate esters, a model with ethylene propylene seals is available. See 

Table

1

.

Table 1. Operating Fluid Compatibility

1. Rotate the reservoir dust cover 1/2 turn. 
2. Fill reservoir approximately 3/4 full with the appropriate fluid. 
3. Rotate dust cover back to cover reservoir.

W

Caution 

To avoid damage to the instrument, check the quality of the operating fluid during 

use. If the fluid becomes discolored, cloudy or particles appear in the reservoir, 

drain and flush the system with clean fluid.

Pressure Range

70 MPa (10 000 psi)

Pressure Connections

M20

Adapters Provided

1/4 NPT Female (quantity 2), 1/2 NPT Female, 1/4 BSP Female, 1/2 BSP 

Female

P5514B-70M Seal Material

Viton®

P5514B-70M-EP Seal 

Material

Ethylene Propylene

Screw Press 

Displacement

20 cm

3

(1.2 in

3

)

Reservoir Volume

75 cm

3

 (4.5 in

3

)

Fine Adjustment 

Displacement

0.06 cm

3

 (0.004 in

3

)

Size (H,W,D)

150 x 330 x 280 mm (6 x 3 x 11 in)

Weight

5 kg (11 lb)

Safety

2014/68/EU, Pressure Equipment Directive (PED) - Article 4, Paragraph 3

Model

O Rings

P5514B-70M

Viton

P5514B-70M-EP

Ethylene Propylene

1.888.610.7664 

[email protected]

Fluke-Direct

.com

Summary of Contents for 5514B-70M

Page 1: ...ump Users Manual April 2021 English 2021 Fluke Corporation All rights reserved Specifications are subject to change without notice All product names are trademarks of their respective companies 1 888...

Page 2: ...ur nearest Fluke authorized service center to obtain return authorization information then send the product to that service center with a description of the difficulty postage and insurance prepaid FO...

Page 3: ...libration 1 Safety Information 1 Specifications 2 Operating Fluid Compatibility 2 DUT Connection 3 Operation 4 Priming 4 Pressure Operation 4 Ancillary Equipment 5 Pointer Remover Punch P5551 5 Master...

Page 4: ...ii P5514B 70M P5514B 70M EP Users Manual 1 888 610 7664 sales GlobalTestSupply com Fluke Direct com...

Page 5: ...is document refer to any pressure measuring instrument such as Transfer Standards Digital Calibrators and Transducers The Fluke 700G and 2700G Digital Test Gauges are recommended for this application...

Page 6: ...trument check the quality of the operating fluid during use If the fluid becomes discolored cloudy or particles appear in the reservoir drain and flush the system with clean fluid Pressure Range 70 MP...

Page 7: ...face of the device to be fitted is clean and undamaged as scratches or dents can form leak paths To connect the DUT 1 Select the appropriate adapter from the adapter set to match the thread on the DUT...

Page 8: ...ssure Operation Follow these steps 1 Turn the screw press in to generate pressure 2 Allow a few moments for the system to stabilize before taking any readings especially after large changes in system...

Page 9: ...can damage the sensitive instruments attached to the system Ancillary Equipment Pointer Remover Punch P5551 To remove and refit the pointer of a pressure gauge use the pointer remover punch instrument...

Page 10: ...s in Table 2 for detailed lists of available ranges and specifications Table 2 Gauges Figure 2 Fluke 700G Model Nominal Specification 700G See Figure 2 0 05 FS 700RG 0 04 Reading 0 01 FS 2700G 0 02 FS...

Reviews: