SC600 Option
Verification
6
6-35
6-51. Edge Amplitude Verification
For the Edge Amplitude verification, connect the Calibrator Mainframe’s SCOPE
connector to the HP 3458A input, using the cable supplied with the Calibrator
Mainframe, the external 50
Ω
termination, and the BNC(f) to Double Banana adapter.
(The 50
Ω
termination is closest to the HP 3458A input.)
1.
For measurements of a 1 kHz signal, set the HP 3458A to DCV, NPLC = .01,
LEVEL 1, TRIG LEVEL, and the DELAY to .0002 for measuring the upper part of
the wave form (i.e. topline), and the DELAY to .0007 for measuring the lower part
of the wave form (i.e. baseline). For measurements of a 10 kHz signal, set the HP
3458A to DCV, NPLC = .001, LEVEL 1, TRIG LEVEL, and the DELAY to .00002
for measuring the topline, and the DELAY to .00007 for measuring the baseline.
2.
Manually lock the HP 3458A to the range that gives the most resolution for the
baseline measurements. Use this same range for the corresponding baseline
measurements at each step. Note that in the EDGE function, the topline is very near
0 V, and the baseline is a negative voltage. See Table 6-24.
3.
For each calibration step, take samples for at least two seconds, using the HP 3458A
MATH functions to enter the average or mean value. See “Setup for SC600 Edge
and Wave Generator Measurements” for more details.
4.
The peak-to-peak value of the wave form is the difference between the topline and
baseline measurements, correcting for the load resistance error. To make this
correction, multiply the readings by (0.5 * (50 + Rload)/Rload), where Rload =
actual feedthrough termination resistance. Record each reading as indicated in Table
6-24.
Table 6-24. Edge Amplification Verification
Calibrator
Mainframe Edge
Output
HP 3458A
Range
Topline
Reading
Baseline
Reading
Peak-to-
Peak
Peak-to-
Peak x
Correction
Tolerance
(
±
V)
100 mV, 1 kHz
100 mV dc
0.0022
1.00V, 1 kHz
1 V dc
0.0202
5 mV, 10 kHz
100 mV dc
0.0003
10 mV, 10 kHz
100 mV dc
0.0004
25 mV, 10 kHz
100 mV dc
0.0007
50 mV, 10 kHz
100 mV dc
0.0012
100 mV, 10 kHz 1 V dc
0.0022
500 mV, 10 kHz 1 V dc
0.0102
1.00 V, 10 kHz
1 V dc
0.0202
2.5 V, 10 kHz
10 V dc
0.0502
6-52. Edge Frequency Verification
This procedure uses the following equipment:
•
PM 6680 Frequency Counter with an ovenized timebase (Option PM 9690 or PM
9691)
Summary of Contents for 5500A
Page 12: ...5500A Service Manual x...
Page 16: ...5500A Service Manual 1 2...
Page 46: ...5500A Service Manual 1 32...
Page 48: ...5500A Service Manual 2 2...
Page 98: ...5500A Service Manual 4 2...
Page 102: ...5500A Service Manual 4 6 om017f eps Figure 4 2 Exploded View of Front Panel Assemblies...
Page 116: ...5500A Service Manual 5 2...
Page 128: ...5500A Service Manual 6 2...
Page 254: ......
Page 255: ......
Page 256: ......
Page 257: ......
Page 258: ......
Page 259: ......
Page 260: ......
Page 261: ......
Page 262: ......
Page 263: ......
Page 264: ......
Page 265: ......
Page 266: ......
Page 267: ......
Page 268: ......
Page 269: ......
Page 270: ......
Page 271: ......
Page 272: ......
Page 273: ......
Page 274: ......
Page 275: ......
Page 276: ......
Page 277: ......
Page 278: ......
Page 279: ......
Page 280: ......
Page 281: ......
Page 282: ......
Page 283: ......
Page 284: ......
Page 285: ......
Page 286: ......
Page 287: ......
Page 288: ...DS2E SL2 DC5V RELAY...
Page 289: ......
Page 290: ......
Page 291: ......
Page 292: ......
Page 293: ......
Page 294: ......
Page 295: ......
Page 296: ......