background image

5500A/EP Enhancement Package
Users Manual

8

Calibrating an Instrument

The DMMs specified in this manual can be adjusted and verified to traceable results, by
following the procedures in this manual. Essentially, there are two major steps to
Calibration, which are:

1.

 

Characterizing the 5500A.

Before you perform any calibration procedure with the 5500A/EP you must
characterize the 5500A’s outputs. The characterization procedure first performs a
zero calibration on the 5500A and an automatic calibration (AUTOCAL) on the
Hewlett Packard 3458A. After the calibration procedures are complete, the Hewlett
Packard 3458A is used to measure the 5500A’s actual output at cardinal points
defined by the UUT procedures. This data is stored in a correction file within
MET/CAL.

The characterization is valid for 24 hours as long as the temperature remains within

±

2

o

C and the humidity does not change more than 

±

10% RH. For convenience, you

can run the characterization unattended overnight, and the characterization will be
valid for the next workday. Or start the characterization procedure at the beginning
of each workday.

2.

 

Adjusting and Verifying the DMM.

Once the characterization is complete, the DMM adjustment and verification
procedures can be performed, provided the MET/CAL procedures have been
installed and compiled on the PC. The MET/CAL program prompts you through the
connection process, allowing for a stabilization period to eliminate thermal emfs in
the system. It also automatically performs a zeroing operation on the 5500A.

Setting up the Equipment

The 5500A/EP is designed to work with the same hardware configuration for both the
characterization of the 5500A, and adjustment and verification of the DMM. The only
difference between the two setups is the use of a Hewlett Packard 3458A for
characterization. Once characterization is complete, you simply disconnect the Hewlett
Packard 3458A and replace it with the DMM you are calibrating, using the same
connections.

Find a location where you can permanently keep the setup. Your location must provide
space for the 5500A calibrator, room to place the switch accessory in front of the 5500A,
and ample room on top or preferably to the left of the 5500A to place the DMM or the
Hewlett Packard 3458A.

Note

When the switch accessory is in operation, high value resistances, which
are highly susceptible to noise pickup, are being sourced. The switch
accessory’s leads contain shielding, and an additional capacitive filter is
available to provide additional filtering. To minimize noise pickup, remain
at a distance of at least 0.5 meters from the switch accessory during
operation.

 Warning

Hazardous Voltage may be present on any Input or Output
terminal of the 5500A/EP switch accessory. To avoid electrical
shock, do not use this accessory unless every lead is
terminated as shown in this manual.

Summary of Contents for 5500A/EP

Page 1: ...602562 August 1996 Rev 3 4 98 1996 1997 1998 Fluke Corporation All rights reserved Printed in U S A All product names are trademarks of their respective companies 5500A EP Enhancement Package Users Ma...

Page 2: ...he warranty period To obtain warranty service contact your nearest Fluke authorized service center or send the product with a description of the difficulty postage and insurance prepaid FOB Destinatio...

Page 3: ...ments 9 Running the Characterization Software 10 Running the Adjustment and Verification Procedures 10 Remote Commands for Operating the Switch Accessory 14 5500A UUT Setup Parameters 14 Format for Ad...

Page 4: ...5500A EP Users Manual ii Switch Accessory Internal Capacitor 27 General Specifications 27...

Page 5: ...iii List of Tables Table Title Page 1 Switch Accessory Lead Functions 6 2 5500A EP 9001 MET CAL Procedures 11 3 Switch Accessory Remote Commands 15...

Page 6: ...5500A EP Users Manual iv...

Page 7: ...ure Title Page 1 5500A EP Enhancement Package Contents 1 2 Battery Eliminators for the 5500 EP Switch Accessory 2 3 Switch Accessory Leads 7 4 Connection for 5500A Characterization 12 5 Connections fo...

Page 8: ...higher 5500A EP Users Manual PN 602562 4 Wire Jumpers Switch Accessory Binding Post Adapters PN 602554 Battery Eliminator RS 232 Cable PN 943738 Calibration Procedure Disk PN 613551 2 3 4 5 1 4 W S H...

Page 9: ...onsists of a male and female connector in a straight through configuration with these connections pin 1 and pin 6 12V pin 2 receive data pin 5 ground Verifying the Battery Eliminator Type for Your Cou...

Page 10: ...n the DOS prompt has the property Close Window on Exit To set that property use the PIF Editor under Program Manager Main to edit the file DOSPRMPT PIF Be certain the box for Close Window on Exit is c...

Page 11: ...ation In the Administrator select File Open then select Traceability 2 To enter traceability information for the Fluke 5500A and the HP 3458A select Edit Traceability to open the Edit Trace Entry Dial...

Page 12: ...ote interface Choose COM5500 as the port Enter the alias as 5500EP Installing the Verification and Adjustment Procedures Place the 5500A EP disk into disk drive A or B and select File Run from the Pro...

Page 13: ...Table 1 Switch Accessory Lead Functions Output Leads Switch Accessory Output DMM or Hewlett Packard 3458A Input 1 A 11A max Current High Input 2 VL1 VL2 V 2W COMP Voltage Low Input 3 VH1 VH2 V 2W COMP...

Page 14: ...L 1 VH 2 VH 1 SL SH HI LO HI LO 11A MAX 5500A EP SWITCH ACCESSORY WARNING Hazardous Voltage may be present on any Input or Output terminal To avoid electrical shock do not use this accessory unless ev...

Page 15: ...C The MET CAL program prompts you through the connection process allowing for a stabilization period to eliminate thermal emfs in the system It also automatically performs a zeroing operation on the 5...

Page 16: ...EEE 488 connector on the 5500A to the IEEE Port 0 on the PC running MET CAL 4 Connect the battery eliminator cable to the switch accessory Plug the switch accessory s battery eliminator into a wall ou...

Page 17: ...pted for the asset number enter the asset number of the EP Switch Accessory and not the 5500A Calibrator Running the Adjustment and Verification Procedures To adjust and verify a DMM make certain the...

Page 18: ...VER IEEE 5500A EP Verifies the Fluke 8842A HP 34401A 1 yr CAL ADJ IEEE 5500A EP Adjusts the Hewlett Packard 34401A HP 34401A 1 yr CAL VER IEEE 5500A EP Verifies the Hewlett Packard 34401A HP 3478A CAL...

Page 19: ...D 1A 250V HI HI LO LO I 200V PK Wall Outlet Fluke 5500A Back Fluke 5500EP HP3458 Front HP3458 Back Yellow White Red Red Black Battery Eliminator Serial 1 From Host IEEE 488 Serial 2 To UUT IEEE 488 A...

Page 20: ...ke 5500EP Fluke 8842 Back Yellow White Red Red Black Battery Eliminator Serial 1 From Host IEEE 488 Serial 2 To UUT IEEE 488 To PC IEEE Port 0 LO HI LO HI Fluke 8842A Front HI HI LO LO 2A INPUT SENSE...

Page 21: ...500A To enter these parameters follow this procedure 1 From the 5500A s Setup menu press the following softkeys to access the REMOTE Setup menu INSTMT SETUP REMOTE SETUP 2 In the Remote Setup menu ver...

Page 22: ...e COMP mode is selected Can also be used when the 5500A is in current mode to carry up to 2A of current from the 5500A 4 Wire Pseudo F9 Relay clusters 2 and 3 are ON Provides a 4 Wire connection to th...

Page 23: ...onnected The 4 Wire Short Circuit should not be used to provide a low ohms short for 2 Wire Ohms calibration Use the 5500A 2 Wire COMP mode instead To provide a reference point for 2 Wire ohms calibra...

Page 24: ...n Measurement ISBN 92 67 10188 9 All specifications are to the cable ends of the 5500A EP switch accessory Part I Specifications Absolute System Uncertainty These specifications show the cardinal poin...

Page 25: ...10 mV 175 ppm 19 mV 105 ppm 30 mV 75 ppm 50 mV 52 ppm 72 5 mV 42 ppm 100 mV 35 ppm 135 mV 32 ppm 190 mV 28 ppm 260 mV 25 ppm 300 mV 24 ppm 0 51V 15 ppm 0 99V to 1 01V and 1 00V to 1 01V 13 ppm 1 9V 12...

Page 26: ...2 2A 200 ppm 2 1 5500A 90 day specification 2 Uncertainty is determined by extrapolation Resistance Cardinal Point Output applies for 2W 4W and pseudo 4W COMP modes 24 hour Uncertainty 2 C 10 rh ppm o...

Page 27: ...m 100 kHz 5600 ppm 28 mV 20 kHz 480 ppm 10 Hz 525 ppm 20 Hz 525 ppm 45 Hz 165 ppm 100 mV 200 Hz 165 ppm 1 kHz 165 ppm 50 kHz 410 ppm 100 kHz 1000 ppm 20 kHz 330 ppm 190 mV 50 kHz 475 ppm 100 kHz 1025...

Page 28: ...pm 1 20 Hz 390 ppm 50 Hz 221 ppm 1 2 8V 20 kHz 280 ppm 50 kHz 425 ppm 100 kHz 950 ppm 3V 1 kHz 175 ppm 50 kHz 415 ppm 3 2V 100 kHz 930 ppm 300 kHz 3400 ppm 10 Hz 370 ppm 10V 1 kHz 150 ppm 50 kHz 410 p...

Page 29: ...Point Output Frequency 24 hour Uncertainty 2 C 10 rh ppm output 45 Hz 920 ppm 10 mA 1 kHz 525 ppm 10 kHz 1850 ppm 30 mA 5 kHz 1200 ppm 1 kHz 550 ppm 100 mA 5 kHz 775 ppm 10 kHz 1950 ppm 190 mA 1 kHz...

Page 30: ...the same loading and application of the environmental conditions that are specified These do not provide error terms for the Hewlett Packard 3458A Repeatability applies for the environmental condition...

Page 31: ...to 109 999 9 2 m 110 to 329 999 9 2 m 330 to 1 09999 k 16 14 m 1 1 k to 3 29999 k 16 14 m 3 3 k to 10 9999 k 18 160 m 11 k to 32 9999 k 18 160 m 33 k to 109 999 k 18 750 m 110 k to 329 999 k 18 750 m...

Page 32: ...z to 100 kHz 200 32 V 100 kHz to 500 kHz 650 70 V 10 Hz to 45 Hz 275 60 V 45 Hz to 10 kHz 50 25 V 0 33V to 3 29999V 10 kHz to 50 kHz 100 50 V 50 kHz to 100 kHz 200 125 V 100 kHz to 500 kHz 450 250 V 1...

Page 33: ...100 nA 1 to 5 kHz 500 100 nA 5 to 10 kHz 1600 10 nA 10 to 20 Hz 500 1000 nA 20 Hz to 45 Hz 300 1000 nA 3 3 mA to 32 999 mA 45 to 1 kHz 250 1000 nA 1 to 5 kHz 500 1000 nA 5 to 10 kHz 1500 1000 nA 10 to...

Page 34: ...certainty and 500 m to the ohms uncertainty Performance not specified in fields 3 V m Relative Humidity Operating 70 to 30 C 60 for 4 wire ohms 10 M Storage 95 noncondensing 1 Altitude Operating 3 050...

Page 35: ...5500A EP Enhancement Package Users Manual 28...

Reviews: