Changing the Thermocouple Type
1. Press
or
until the display shows
TYPE
.
2. Press
to display the thermocouple type choices.
The currently selected thermocouple blinks.
3. Press
or
until the thermocouple you want
appears on the display.
4. Press
to store the thermocouple type in memory.
Changing the Offset
You can adjust the thermometer's readings to compensate
for the errors of a specific thermocouple. See "Using the
Offset to Adjust for Probe Errors." The allowable
adjustment range is
±
5.0
o
C or K, and
±
9.0
o
F.
Model 54:
You can store individual offsets for T1 and T2.
1. Press
or
until the display shows
and
T1
or
T2
.
2. Press
to indicate that you want to change the
offset setting.
The temperature measurement plus the offset
appears in the primary display. The offset appears in
the secondary display.
3. Press
or
until the primary display shows the
correct reading.
4. Press
to store the offset setting in memory.
Remember to reset the offset to 0.0 when it is no longer
needed. The offset automatically resets to 0.0 when you
change the thermocouple type.
Summary of Contents for 53IIB