background image

 

Scope Calibration Option 

 

Calibrating the Voltage Amplitude on an Oscilloscope 

Calibrating the Voltage Amplitude on an Oscilloscope 

The oscilloscope voltage gain is calibrated by applying a low frequency square wave 

signal and adjusting its gain to meet the height specified for different voltage levels, 

designated by the graticule line divisions on the oscilloscope. The signal is applied from 

the 5080A in Volt mode. The specific voltages that you should use for calibration, and 

the graticule line divisions that need to be matched, vary for different oscilloscopes and 

are specified in your oscilloscope’s service manual. 

The Volt Function 

The Voltage gain is calibrated using the Volt function. This function is accessed through 

the Volt menu, which appears when you start the SCOPE option, or when you press the 

softkey under MODE to scroll through the oscilloscope calibration menus. 

 

gjm003.eps 

Each menu item is described below: 

 

OUTPUT @ SCOPE

  Indicates the location of the signal output.  If the signal does 

not appear on the oscilloscope, press 

. To disconnect the signal, press 

 

DC -> AC

 or 

DC <- AC

 Toggles between a dc and ac signal. Pressing the softkey 

from the ac signal produces the dc equivalent output. 

 

SCOPE Z

  Toggles the calibrator’s output impedance setting between 1 M

Ω

 and  

50 

Ω

 

V/DIV MENU

  Opens the voltage scaling menu, which lets you select the scale of 

the signal in volts per division. This menu is described below in detail, under “The 

V/DIV Menu.” 

 

MODE

  Indicates you are in Volt mode. Use the softkey to change modes and open 

the corresponding menus for the other three oscilloscope calibration modes. 

The V/DIV Menu 

The V/DIV menu, shown below, sets the number of volts denoted by each division on the 

oscilloscope. This menu provides alternative methods for changing the output amplitude 

that may be more convenient for certain oscilloscope applications. To access the V/DIV 

menu, press V/DIV from the Volt menu. 

riešenia na presné meranie

http://www.elso.sk

Summary of Contents for 5080A/SC

Page 1: ...rights reserved Printed in USA Specifications are subject to change without notice All product names are trademarks of their respective companies 5080A SC Oscilloscope Calibration Option Users Manual...

Page 2: ...zed service center to obtain return authorization information then send the product to that service center with a description of the difficulty postage and insurance prepaid FOB Destination Fluke assu...

Page 3: ...ltage Amplitude on an Oscilloscope 7 The Volt Function 7 The V DIV Menu 7 Shortcuts for Setting the Voltage Amplitude 8 Amplitude Calibration Procedure for an Oscilloscope 8 Calibrating the Pulse and...

Page 4: ...5080A SC Users Manual ii rie enia na presn meranie http www elso sk...

Page 5: ...Load 21 5 Voltage Function Verification DC Voltage into a 50 Load 22 6 Voltage Function Verification DC Voltage into a 1 M Load 22 7 Edge Function Verification 23 8 Leveled Sine Wave Function Verific...

Page 6: ...5080A SC Users Manual iv rie enia na presn meranie http www elso sk...

Page 7: ...v List of Figures Figure Title Page 1 Oscilloscope Connection Channel and External Trigger 4 rie enia na presn meranie http www elso sk...

Page 8: ...5080A SC Users Manual vi rie enia na presn meranie http www elso sk...

Page 9: ...nction This calibration procedure is similar to the one for verifying the vertical deflection characteristics except that it checks the horizontal axis The menus that implement these functions also in...

Page 10: ...200 m 40 000 ft maximum Detailed Specifications Voltage Function DC Signal Square Wave Signal Voltage Function Into 50 Into 1 M Into 50 Into 1 M Amplitude Characteristics Range 0 V to 2 2 V 0 V to 33...

Page 11: ...ess relative to 50 kHz Not applicable 1 5 of output 200 V 2 0 of output 200 V Short term Stability 1 1 Frequency Characteristics Resolution 10 Hz 10 kHz 2 10 kHz Specification 1 Year tcal 5 C 5 ppm 5...

Page 12: ...e the external trigger attach the TRIG OUT connector on the 5080A to the external trigger connection on your oscilloscope To use the external trigger and view its signal with the calibration signal at...

Page 13: ...Since oscilloscope calibration requires many adjustments of the output signal the three available methods for changing these settings for oscilloscope calibration are summarized below These methods p...

Page 14: ...ps 5 Press to remove the cursor from the Output Display and save the new value as the reference value Note If you attempt to use the rotary knob to adjust a value to an amount that is invalid for the...

Page 15: ...tion menus gjm003 eps Each menu item is described below OUTPUT SCOPE Indicates the location of the signal output If the signal does not appear on the oscilloscope press To disconnect the signal press...

Page 16: ...Shortcuts for Setting the Voltage Amplitude The and keys step the voltages through cardinal point values of an oscilloscope in a 1 2 5 step sequence For example if the voltage is 40 mV then pressing i...

Page 17: ...mount specified for the calibration settings for your oscilloscope This example shows the gain at 30 mV to be 6 divisions at 5 mV per division gl006i bmp 4 Change the voltage to the next value recomme...

Page 18: ...Indicates you are in Edge mode Use the softkey to change modes and open the corresponding menus for the other three oscilloscope calibration modes Pulse Response Calibration Procedure for an Oscillos...

Page 19: ...oscilloscope exhibits the proper rise time and pulse aberration characteristics 7 Remove the input signal by pressing The Leveled Sine Wave Function The Leveled Sine Wave Levsine function uses a level...

Page 20: ...ortcuts for Setting the Frequency and Voltage Three options are available for controlling the sine wave settings SET TO LAST F toggles between the last frequency used and the reference frequency of 50...

Page 21: ...sweep rate lets you watch the frequency change very slowly After a fast sweep you may want to pinpoint a certain frequency with a slow sweep over a subset of your previous frequency range RANGE The so...

Page 22: ...the Sweep menu appears on the Control Display as shown below gl031i eps 4 You can let the signal sweep through the entire range or you can halt the sweep if you need to record the frequency at a certa...

Page 23: ...y six divisions To fine tune the voltage press to bring a cursor into the Output Display move the cursor with the key and turn the rotary knob to adjust the value See Fine Tuning Values earlier in thi...

Page 24: ...is available through the Marker menu lets you calibrate the timing response of your oscilloscope To access the Marker menu press the softkey under MODE until marker appears gjm008 eps Each option in t...

Page 25: ...point which is 500 s Time Base Marker Calibration Procedure for an Oscilloscope This sample procedure uses the Time Marker function to check the horizontal deflection time base of your oscilloscope S...

Page 26: ...the 5080A Operators Manual Overlapped Commands Commands that require additional time to execute are called overlapped commands because they can overlap the next command before completing execution To...

Page 27: ...scope Marker mode Programs the period to 1 ms output at the SCOPE BNC standby if from OFF or previously in standby Example SCOPE VOLT OUT 2 V 0 Hz dc voltage 2 V SCOPE VOLT OUT 4 V 1 kHz ac voltage 4...

Page 28: ...points The verification test points are provided here as a guide when re verification is desired Table 3 Voltage Function Verification AC Voltage into a 1 M ohm Load Nominal Value p p Frequency Measur...

Page 29: ...cation AC Voltage into a 50 Load Nominal Value p p Frequency Measured Value p p Deviation mV 1 Year Spec mV 5 0 mV 45 Hz 0 2175 5 0 mV 100 Hz 0 2175 5 0 mV 1 kHz 0 2175 10 0 mV 100 Hz 0 235 10 0 mV 1...

Page 30: ...0 2875 55 0 mV 0 3925 55 0 mV 0 3925 100 0 mV 0 55 100 0 mV 0 55 220 0 mV 0 97 220 0 mV 0 97 250 0 mV 1 075 250 0 mV 1 075 550 0 mV 2 125 550 0 mV 2 125 700 0 mV 2 65 700 0 mV 2 65 2 2 V 7 9 2 2 V 7...

Page 31: ...1 95 500 0 mV 1 95 3 3 V 11 75 3 3 V 11 75 4 0 V 14 2 4 0 V 14 2 33 0 V 115 7 33 0 V 115 7 Table 7 Edge Function Verification Nominal Value p p Frequency Pulse Response Time nS 1 Year Spec ns 250 0 mV...

Page 32: ...le 9 Leveled Sine Wave Function Verification Flatness Nominal Value p p Frequency Measured Value p p Deviation mV 1 Year Spec mV 5 0 mV 500 kHZ 0 18 5 0 mV 1 MHz 0 18 5 0 mV 1 MHz 0 18 5 0 mV 2 MHz 0...

Page 33: ...0 40 0 mV 5 MHz 0 70 40 0 mV 10 MHz 0 70 40 0 mV 20 MHz 0 70 40 0 mV 50 MHz 0 70 40 0 mV 100 MHz 0 70 40 0 mV 125 MHz 0 90 40 0 mV 160 MHz 0 90 40 0 mV 200 MHz 0 90 40 0 mV 220 MHz 0 90 40 0 mV 235 MH...

Page 34: ...10 400 0 mV 125 MHz 8 10 400 0 mV 160 MHz 8 10 400 0 mV 200 MHz 8 10 1 3 V 500 kHZ 19 60 1 3 V 1 MHz 19 60 1 3 V 1 MHz 19 60 1 3 V 2 MHz 19 60 1 3 V 5 MHz 19 60 1 3 V 10 MHz 19 60 1 3 V 20 MHz 19 60...

Page 35: ...V 10 MHz 0 0003 MHz 1 3 V 200 MHz 0 0050 MHz Table 11 Marker Generator Function Verification Nominal Interval Measured Interval Deviation 1 Year Spec 5 s 25 13 ms 2 00 s 4 05 ms 1 s 1 03 ms 500 00 ms...

Page 36: ...enerator Function Verification cont Nominal Interval Measured Interval Deviation 1 Year Spec 100 00 ns 2 500 ps 50 00 ns 1 250 ps 20 00 ns 0 500 ps 10 00 ns 0 250 ps 5 00 ns 0 125 ps 2 00 ns 0 050 ps...

Reviews: