Appendix
Operation Instructions
B
B-9
B.4.6 Annunciators, Indicators, and Overrange Conditions
Table B-1 specifies the annunciators that may appear on the top line of the display.
Table B-1. Annunciators
Annunciator Condition
??Bias
No bias measurement possible since the
electrometer is not ready
ADC
Air density correction in use (only applicable to
radiological measurements)
ALC
Automatic leakage compensation enabled
LoBat
Low Battery voltage
LoBias
Low Bias voltage
HiBias High
Bias
voltage
OvAmps
Overranged current
Rst Resetting
electrometer
Tmr Timer
in
use
WAIT Bias Not Ready
The dosimeter is setting the bias
WAIT System Not Rdy
The dosimeter is resetting the electrometer
(only applicable when automatic leakage
compensation is enabled)
Figure B-8 illustrates the locations of some annunciators. The battery, bias, and overranged current
annunciators appear at the same location, and the instrument alternates the displaying of each
annunciator when more than one is applicable.
Figure B-8.
Annunciator Positions
Table B-2 lists the indicators the instrument uses to signal that a measurement quantity has not been
compensated for leakage or is not available due to the specified condition.
Table B-2. Measurement Indicators and Conditions
Indicator Measurement
Condition
?
(left of value)
All
Non-leakage
1
Automatic leakage compensation is enabled, but the exposure
started before a leakage measurement was available.
-.---
Leakage
Automatic leakage compensation is enabled, but the exposure
started before a leakage measurement was available.
INVALID
Effective Time
The calculated effective time is longer than the actual
exposure. This condition occurs when a large amount of
charge is injected at the beginning of the exposure.
NA
Leakage
The leakage measurement is not available since automatic
leakage compensation is disabled.
OVER Non-leakage
1
The quantity or one of its dependents has exceeded the
maximum value. Refer to Table B-3 for more information.
1
All non-leakage measurements (e.g., charge, current, dose, rate, effective time, etc.)
3 9 . 9 9 5 n C
5 9 . 8 9 1 s
6 5 9 . 8 p A
A D C L o B i a s T m r A L C
Summary of Contents for 35040 Operators
Page 6: ...Blank page...
Page 70: ...35040 Operators Manual Blank Page...
Page 72: ...35040 Operators Manual A 2...
Page 73: ...Appendix Sample Customization Reports A A 3...
Page 74: ...35040 Operators Manual A 4...
Page 75: ...Appendix Sample Customization Reports A A 5...
Page 76: ...35040 Operators Manual Blank page...
Page 93: ......