Clamp
Meter
How to Clean the Product
9
Note
The Measurement Category (CAT) and voltage rating of combinations of test
probes, test probe accessories, current clamp accessories, and the Product is the
LOWEST rating of individual components.
How to Clean the Product
Regularly wipe the case with a damp cloth and weak detergent.
W
Caution
To prevent damage to the Product, do not use abrasives or solvents to clean
the Product case.
To clean the Product jaw:
1.
Examine the jaw mating surface to make sure it is clean. If there is unwanted material
(including rust), jaw closure will not be correct and there will be measurement errors.
2.
Open the jaws and clean the clamp metal ends with a lightly oiled cloth.