Fluke 3014-PR Instruction Sheet Download Page 4

3014-PR English Instruction Sheet 

Page 

 

 

Overload 

50 A for Continuous 

Frequency Range 

40 to 10 kHz (See Figure 2) 

Crest Factor 

3 @ 10 A rms with an error (due to CF) of 

3 %  

Working Voltage 

600 V rms 

Common Mode Voltage 

30 V rms 

Influence of Adjacent 

Conductor 

<15 mA/A at 50 Hz 

Influence of dc on ac signal  dc current <20 A, <5 % 

Influence of Conductor 

Position in Jaw 

0.5 % of mV output @ 50/60 Hz 

Influence of Frequency 

40 Hz to 1 kHz: 0.5 % of mV output  

1 kHz to 10 kHz: 2 % of mV output 

 

Mechanical Specifications 

Dimensions 

5.47 x 2.00 x 1.18 in. (139 x 51 x 30 mm) 

Weight 

180 g (6.5 oz) 

Jaw Opening 

0.83 in. (21 mm) 

Maximum Conductor Size 

  Cable 

  Bus Bar 

 

0.78 in. (20 mm)  

0.78 x 0.19 in. (20 x 5 mm) 

Case Protection 

IP 40 (IEC 60529) 

Drop Test 

1.0 m six sides to oak floor 

Vibration MIL-PRF-28800F 

Polycarbonate Material 

  Jaws 

  Case 

 

Polycarbonate Red UL 94 V0  

Polycarbonate ABS Gray: UL 94 V0 

Opening Operations - Life 

>50,000 

Output 

 

1 ft round 2 conductor cable w/4-pin Lemo 

connector  

 

Summary of Contents for 3014-PR

Page 1: ...s the latest safety and performance standards Symbols The table below lists the symbols used on the Clamp and or in this instruction sheet Symbol Description Do not dispose of this product as unsorted municipal waste Go to Fluke s website for recycling information W Risk of Danger Important Information See manual X Hazardous Voltage Risk of electric shock T Double insulation Application around and...

Page 2: ...the operating instructions otherwise the safety features may not protect you Adhere to local and national safety codes Individual protective equipment must be used to prevent the shock and arc blast injury where hazardous live conductors are exposed Before each use inspect the Clamp Look for cracks or missing portions of the housing or output cable insulation Also look for loose or weakened compon...

Page 3: ...Clean the Clamp periodically by wiping it with a damp cloth and detergent Do not use abrasive cleaners or solvents Do not immerse the Clamp in liquids Specifications Electrical Specifications Operating Range 40 A Measurement Range 0 1 to 40 A Transformation Ratio Voltage output Output Signal 50 mV A 2 V 40 A Range Accuracy Phase Shift 0 1 to 1 A 2 of Reading 1 mV 1 to 5 A 1 of Reading 1 mV 3 5 to ...

Page 4: ...t 50 60 Hz Influence of Frequency 40 Hz to 1 kHz 0 5 of mV output 1 kHz to 10 kHz 2 of mV output Mechanical Specifications Dimensions 5 47 x 2 00 x 1 18 in 139 x 51 x 30 mm Weight 180 g 6 5 oz Jaw Opening 0 83 in 21 mm Maximum Conductor Size Cable Bus Bar 0 78 in 20 mm 0 78 x 0 19 in 20 x 5 mm Case Protection IP 40 IEC 60529 Drop Test 1 0 m six sides to oak floor Vibration MIL PRF 28800F Polycarbo...

Page 5: ...hout roll off above 35 C Influence of Temperature 150 ppm K or 0 15 of mV output per 10 K Influence of Humidity 10 to 90 RH 0 2 of mV output per 10 K 18 C to 28 C Safety Specifications Electrical Conforms to EN IEC 61010 and 61010 2 032 600 V Cat III Pollution Degree 2 Electromagnetic Compatibility EN 50081 1 Class B EN 50082 2 Electrostatic discharge IEC 61000 4 2 Radiated field IEC 61000 4 3 Fas...

Page 6: ... operation or handling Resellers are not authorized to extend any other warranty on Fluke s behalf To obtain service during the warranty period contact your nearest Fluke authorized service center to obtain return authorization information then send the product to that Service Center with a description of the problem THIS WARRANTY IS YOUR ONLY REMEDY NO OTHER WARRANTIES SUCH AS FITNESS FOR A PARTI...

Reviews: