Appendices
Specifications
A
A-15
Table A-19. PAI Module Thermocouple Specifications
Accuracy
+
°
C
Thermocouple Resolution
18
°
C to 28
°
C -10
°
C to 60
°
C
ITS90
°
C
90 Day
1 Year
1 Year
Type Temperature
°
C Slow
Slow
Fast
Slow
Fast
E
-100 to –25
.03
0.45
0.50
0.80
0.60
0.80
-25
to
20 .02
0.35
0.40
0.60
0.50
0.70
20
to
600 .02
0.30
0.40
0.60
0.50
0.80
600 to 1000
.02
0.40
0.50
0.70
0.90
1.00
J
-100 to 80
.03
0.45
0.50
0.80
0.60
0.80
80
to
230 .02
0.35
0.50
0.70
0.60
0.80
230 to 760
.02
0.40
0.50
0.70
0.80
0.90
K
-100 to –25
.04
0.55
0.60
0.90
0.70
1.00
-25
to
120 .03
0.40
0.50
0.80
0.60
0.90
120 to 800
.03
0.50
0.65
0.90
1.00
1.20
800 to 1372
.03
0.70
1.00
1.30
1.60
1.90
L
-100 to 100
.04
0.90
1.00
2.00
1.30
2.00
100 to 800
.04
0.50
0.90
1.40
1.20
1.70
800 to 900
.03
0.50
0.70
1.10
1.30
1.50
N
-100 to –25
.05
0.65
0.75
1.20
0.80
1.30
-25
to
120 .05
0.55
0.60
1.00
0.70
1.10
120 to 1000
.04
0.45
0.60
0.90
1.00
1.20
1000 to 1300
.03
0.55
0.75
1.00
1.20
1.50
R
250 to 600
0.1
0.90
1.00
2.10
1.20
2.20
600 to 1500
0.1
0.80
0.90
1.80
1.30
2.00
1500 to 1767
0.1
0.85
0.85
1.90
1.70
2.50
S
250 to 1000
0.1
0.95
1.10
2.30
1.30
2.40
1000 to 1400
0.1
0.80
1.00
1.90
1.40
2.30
1400 to 1767
0.1
1.00
1.30
2.20
1.80
2.80
Summary of Contents for 2680A
Page 8: ......
Page 14: ...2680A 2686A Users Manual vi ...
Page 20: ...2680A 2686A Users Manual 1 2 ...
Page 98: ...2680A 2686A Users Manual 2 62 ...
Page 196: ......
Page 228: ...2680A 2686A Users Manual A 32 ...
Page 236: ...2680A 2686A Users Manual C 4 ...
Page 242: ...2680A 2686A Users Manual D 6 ...
Page 274: ...2680A 2686A Users Manual 6 ...