Manual Supplement
1652C/1653B/1654B Calibration
2 4/17
Change #2
On page 15, add the following to Table 2:
Touch plate
FLUKE-165X-8006,TOUCH PLATE
Fluke P/N 1884378
On page 25, replace steps 1 through 8 with:
1.
Set the UUT rotary switch to
LOOP Z
I
NO TRIP
function.
2.
Using
<
, set the UUT for L-PE test.
3. Connect test leads to the UUT L-PE-N input jacks. Short the far end of the test leads together with a
Zero Adapter (Fluke P/N 3301338). See Figure 8.
4. Press and hold
for approximately three seconds until the UUT ZERO
=
annunciator appears
.
On page 27, replace steps 1 through 7 with:
1.
Set the UUT rotary switch to LOOP Z
I
HI CURRENT function.
2. Connect test leads to the UUT L-PE-N input jacks. Short the far end of the test leads together with a
Zero Adapter (Fluke P/N 3301338). See Figure 8.
3. Press and hold
for approximately three seconds until the UUT ZERO
=
annunciator appears
.
On page 28, replace step 7 with:
7. Using the 5320A cursor keys or keypad, set the 5320A Nominal R Value to the nearest value to
those in Table 6 that the 5320A will source. For Steps 1 and 8, take the first value that is
t
0.2 .
On page 28, replace Table 6 with:
Table 6. Loop ZI Hi Current Accuracy Test
Step
UUT
Function
UUT
<
Setting
5320A
Nominal
R Value
5320A
Residual
Impedance
Correction
Type
Recorded
5320A
Displayed
Output Z
Value
UUT
Accuracy
Spec.
±(% + dig)
UUT
Res.
Calculated UUT
Display Limits
Lower
Upper
1
Loop ZI Hi
Current
L-PE
0.2
Comp
+
(2 % + 4)
0.01
2
1.8
3
18
4
50
+2 %
0.1
5
180
6
500
+6 %
1
7
1800
8
L-N
0.2
+(2 % + 4)
0.01
9
1.8
10
18
11
50
+
2 %
0.1
12
180
13
500
+
6 %
1
14
1800