![Fluke 123B/NA User Manual Download Page 82](http://html1.mh-extra.com/html/fluke/123b-na/123b-na_user-manual_2309470082.webp)
123B/124B/125B
Users Manual
74
Accuracy
125B, 124B
@ 1 Hz to 1 MHz ........................................
±
(0.5 % + 2 counts)
@ 1 to 10 MHz ............................................
±
(1.0 % + 2 counts)
@ 10 to 70 MHz ..........................................
±
(2.5 % + 2 counts)
123B
@ 1 Hz to 1 MHz ........................................
±
(0.5 % + 2 counts)
@ 1 to 10 MHz ............................................
±
(1.0 % + 2 counts)
@ 10 to 50 MHz ..........................................
±
(2.5 % + 2 counts)
(50 MHz in Autorange)
Full Scale Reading .............................................. 10 000 counts
RPM
Max reading ........................................................ 50.00 kRPM
Accuracy .............................................................
±
(0.5 % + 2 counts)
Duty Cycle (PULSE)
Range ................................................................. 2 % to 98 %
Frequency Range in Continuous Autoset ........... 15 Hz (1 Hz) to 30 MHz
Accuracy (Logic or Pulse waveforms)
@ 1 Hz to 1 MHz ............................................
±
(0.5 % + 2 counts)
@ 1 MHz to 10 MHz .......................................
±
(1.0 % + 2 counts)
Pulse Width (PULSE)
Frequency Range in Continuous Autoset ........... 15 Hz (1 Hz) to 30 MHz
Accuracy (Logic or Pulse waveforms)
@ 1 Hz to 1 MHz ............................................
±
(0.5 % + 2 counts)
@ 1 MHz to 10 MHz .......................................
±
(1.0 % + 2 counts)
Full Scale Reading .......................................... 1000 counts
1.888.610.7664