2.6 Electrical  interface 

The Local alarm (1060-024) requires power connections as follows: (See appendix A for 
the 1060-13 and 1060-024 drawings) 

Terminate the input power to TB2-1(L1) of the Local alarm assembly at the end of 
this installation .   
 

  

1.

 

Earth ground to chassis lug items 17 and 18  in view A-A of  the local alarm 
assembly. 

2.

 

The L2 neutral connection (white) to TB1-8. 

3.

 

Connect the 75 foot power interface (male end) cables (1060-022 ) to both J1 
and J2 . They are interchangeable. 

4.

 

Remove the dust cap and connect the 1060ds-24 ten foot cable assembly to  

     J3 and connect the free end to the I/O connector on the 1060am-nm-LR .                     
 
 

The Remote alarm (1060-034) require power connections as follows: .(See appendix  A 
for the 1060-023 and 1060-034 drawings) 
 

1.

 

Earth ground to chassis lug items 21 and 22 in view A-A  of  the remote alarm 
assembly. 

2.

 

The L2 neutral (L2) connection (white) to TB1-8. 

3.

 

Connect either female end of the power interface cable to P1 on the remote 
alarm  chassis . 

 

The Flasher alarm (1060-026) requires power connections as follows: 
(see appendix A for 1060-033 and 1060-026 drawings) 
 

1.

 

Earth ground to chassis lug of the flasher alarm assembly. 

2.

 

Neutral (L2) connection to ac-flasher terminal 1.  

3.

 

Connect the remaining female end of the power interface cable to P1. 

 
 
   

 

 

 

 

 

Summary of Contents for 1060AM-NM-LRVAR Operators

Page 1: ...cal Remote and Flasher Alarms Operators Manual March 2005 Manual No 1060AMNMLRVAR 1 Rev 1 2004 2005 Fluke Corporation All rights reserved Printed in U S A All product names are trademarks of their res...

Page 2: ...Fluke Biomedical Radiation Management Services 6045 Cochran Road Cleveland Ohio 44139 440 498 2564 www flukebiomedical com rms...

Page 3: ...General Description 2 1 2 2 Specifications 2 1 2 3 Receiving Inspection 2 2 2 4 Storage 2 2 2 5 Installation 2 3 2 6 Electrical Interface 2 4 2 7 Operation 2 5 Section 3 Maintenance and Cleaning 3 1...

Page 4: ...Blank Page...

Page 5: ...S is at the user s own risk and is sufficient cause to terminate the warranty agreement between Fluke Biomedical Radiation Management Services and the customer Adequate warnings are included in this m...

Page 6: ...1060AM NM LRVAR Operators Manual Blank Page...

Page 7: ...f normally closed contacts J3 is a male connector with a dust cover that interfaces with the 1060am nm LR A inch conduit fitting is provided for the hot L1 neutral L2 and earth ground connections The...

Page 8: ...d in ANSI N45 2 2 1972 Section 6 1 2 2 The storage area shall comply with ANSI N45 2 2 1972 Section 6 2 Storage Area Paragraphs 6 2 1 through 6 2 5 Housekeeping shall conform to ANSI N45 2 3 1972 Leve...

Page 9: ...hr A field between 10mr hr and a 100mr hr is good starting point Contact your radiation safety officer for your company requirements If readings from a device s position indicates that the radiation r...

Page 10: ...the 1060ds 24 ten foot cable assembly to J3 and connect the free end to the I O connector on the 1060am nm LR The Remote alarm 1060 034 require power connections as follows See appendix A for the 1060...

Page 11: ...blinking Connect the free end of the I O 1060ds 24 cable to the model 1060AM NM LR I O connector Connect the 1060AM NM 110 12 volt 500 mA power cube connector to the 1060AM NM LR power input The red...

Page 12: ...1060AM NM LRVAR Operators Manual Blank Page...

Page 13: ...o not touch glass with bare fingers Grasp glass with a soft clean cloth or with packaging supplied with the replacement lamp 2 Gently push down on the lamp twist and pull up to remove the lamp from th...

Page 14: ...y Ensure that the acrylic sign is completely dry before assembling the unit 3 1 Troubleshooting Extreme care must be used when troubleshooting a system that has power applied All standard troubleshoot...

Page 15: ...Assembly 1060 010 Alarm PCB Parts List 1060 013 Local Alarm Schematic 1060 024 Local Alarm Assembly 1060 024 Local Alarm Parts List 1060 023 Remote Alarm Schematic 1060 034 Remote Alarm Assembly 1060...

Page 16: ...Fluke Biomedical Radiation Management Services 6045 Cochran Road Cleveland Ohio 44139 440 498 2564 www flukebiomedical com rms...

Reviews: