Operation
Procedure for Removing Trapped Air from System
3
3-3
1.
Open HP valve (1) and LP valve (2) fully, (counter-clockwise).
2.
Open reservoir valve (3), (approximately 4 turns counter-clockwise).
3.
Turn capstan fully out (counter-clockwise), and wait for approximately 60 seconds.
glf05.bmp
Figure 3-4. Turn Capstan Fully Clockwise
4.
Turn capstan fully in (clockwise), and wait for approximately 15 seconds.
5.
Trapped air from the system will appear as bubbles in the reservoir.
6.
Repeat the procedure until no more air appears in the reservoir.