Fluke Calibration P5514B-70M User Manual Download Page 7

Hydraulic Comparison Test Pump

DUT Connection

3

DUT Connection

W

Caution

To prevent Product damage or damage to a gauge: 
• To prevent incorrect sealing that creates an unsafe condition, do not use thread-

sealing tape or any other sealing method on the DUT or the adapters. The Gauge 

Adapter sealing system can be hand-tight sealed up to 70 MPa (10500 psi). 

Wrenches or similar tools are not required. Over tightening can cause damage to 

threads or sealing faces. 

• Before connection, ensure that there are O-rings fitted to the test port and gauge 

adapter. 

• Check that the sealing face of the device to be fitted is clean and undamaged, as 

scratches or dents can form leak-paths. 

To connect the DUT:
1. Select the appropriate adapter from the adapter set to match the thread on the DUT.
2. Screw the adapter fully onto the DUT so that the bottom face of the DUT seals on the O-

ring inside the adapter. Hand-tight is sufficient.

3. Turn the adapter collar counter-clockwise until the adapter seals on the O-ring in the 

mounting post to mount the DUT/adapter assembly to the test port. Hand-tight is 

sufficient.

4. To adjust the position of the DUT, hold the adapter, and turn the collar clockwise 1/4-turn.
5. Position the DUT to face the desired direction, and turn the collar counter-clockwise to re-

seal.

Summary of Contents for P5514B-70M

Page 1: ...P Hydraulic Comparison Test Pump Users Manual April 2021 English 2021 Fluke Corporation All rights reserved Specifications are subject to change without notice All product names are trademarks of thei...

Page 2: ...n warranty service contact your nearest Fluke authorized service center to obtain return authorization information then send the product to that service center with a description of the difficulty pos...

Page 3: ...ion 1 How to Contact Fluke Calibration 1 Safety Information 1 Specifications 2 Operating Fluid Compatibility 2 DUT Connection 3 Operation 4 Priming 4 Pressure Operation 4 Ancillary Equipment 5 Pointer...

Page 4: ...ii P5514B 70M P5514B 70M EP Users Manual...

Page 5: ...st Gauge used The Master Test Gauge must be regularly calibrated to ensure accuracy is maintained How to Contact Fluke Calibration Fluke Corporation operates worldwide For local contact information go...

Page 6: ...ion To avoid damage to the instrument check the quality of the operating fluid during use If the fluid becomes discolored cloudy or particles appear in the reservoir drain and flush the system with cl...

Page 7: ...apter Check that the sealing face of the device to be fitted is clean and undamaged as scratches or dents can form leak paths To connect the DUT 1 Select the appropriate adapter from the adapter set t...

Page 8: ...during priming operation Pressure Operation Follow these steps 1 Turn the screw press in to generate pressure 2 Allow a few moments for the system to stabilize before taking any readings especially a...

Page 9: ...will shock the system which can damage the sensitive instruments attached to the system Ancillary Equipment Pointer Remover Punch P5551 To remove and refit the pointer of a pressure gauge use the poi...

Page 10: ...ruction manuals for the gauges in Table 2 for detailed lists of available ranges and specifications Table 2 Gauges Figure 2 Fluke 700G Model Nominal Specification 700G See Figure 2 0 05 FS 700RG 0 04...

Reviews: