Maintenance & Servicing
Draining PCU Fluid
5
5-3
1
2
5
3
4
gmg11.eps
Figure 5-1. PCU Assembly
Table 5-1. PCU Assembly - Parts List
Item Description
Part
1 Weight
Carrier
SPEC
2 Set
Screw
3910313
3 Cylinder
SPEC
4 Piston
SPEC
5 Cylinder
Retainer 3919216
Draining PCU Fluid
Note
In order to prevent contamination of the pneumatic system, the PCU
lubricating fluid must be drained from the PCU housing before
commencing any maintenance procedures that involve moving the
instrument from the horizontal position. This includes packing the
instrument for storage or shipment.
1.
Depressurize the system by slowly opening the exhaust valve.
2.
Disconnect any DUT from the test port.
3.
Place absorbent cloth or tissue paper around the base of the PCU housing.
4.
Unscrew drain plug at the lower, rear of the housing to release fluid.
5.
To increase flow rate, unscrew filler plug at the upper left hand side of the housing.
Summary of Contents for P3031
Page 6: ...P3031 P3032 Users Manual iv...
Page 8: ...P3031 P3032 Users Manual vi...
Page 18: ...P3031 P3032 Users Manual 2 6...
Page 22: ...P3031 P3032 Users Manual 4 2...
Page 30: ...P3031 P3032 Users Manual 5 8...
Page 32: ...P3031 P3032 Users Manual 6 2...
Page 34: ...P3031 P3032 Users Manual 7 2...