Fluke Calibration E-DWT 3622351 Instruction Sheet Download Page 2

 

6.  Loosen the hex nut using a 3/8 inch socket. If 

recessed area is too small to allow the socket to 
fit, use needle nose pliers.       

7.  Locate the internal 12V cable that connects to the 

RPM4 (see Figure 4). 

 

Figure 4:  E-DWT internal view detail: 12V grey power 

cable

 

8.  Loosen the knurled nut and disconnect the cable 

from the RPM4 (see Figure 5). 

 

Figure 5:  E-DWT internal view detail: knurled nut to 

RPM4

 

 

 

 

 

9.  Install the new cable.  

10.  Connect the new cable (knurled nut end) to the RPM4 

(see Figure 5). Secure it by manually tightening 
finger tight.   

11.  Remove the hex nut and washer from the heat-

shrinked connector, (see Figure 4) insert the 
connector through the E-DWT housing, and secure 
it to the outside of the housing using the washer and 
hex nut that was just removed. The washer should 
be installed on the outside of the housing directly 
under the nut. It is recommended to use Loctite 
222MS or similar on the nut to help ensure it does 
not come loose. 

12.  Reposition the E-DWT on the table in its normal 

upright position. 

13.  Replace the variable volume handle that was 

removed (see Figure 2). 

14.  Connect Adapter to end of power supply cable.  

15.  Connect new power supply and open vent on 

reservoir for normal operation (see Figure 1). 

16.  Test functionality to ensure there are no problems 

with the cable installation or power supply. 

17.  Dispose of old cable and old power supply. 

 