5522A
Getting Started
12
Relative Humidity
Operating ............................................................ <80 % to 30
°
C, <70 % to 40
°
C, <40 % to 50
°
C
Storage ............................................................... <95 %, non-condensing. After long periods of storage at high humidity,
a drying-out period (with power on) of at least one week may be
required.
Altitude
Operating ............................................................ 3,050 m (10,000 ft) maximum
Non-operating ..................................................... 12,200 m (40,000 ft) maximum
Safety
..................................................................... Complies with EN/IEC 61010-1:2001, CAN/CSA-C22.2 No.
61010-1-04, ANSI/UL 61010-1:2004;
Output Terminal Electrical Overload Protection
Provides reverse-power protection, immediate output disconnection,
and/or fuse protection on the output terminals for all functions. This
protection is for applied external voltages up to
±
300 V peak.
Analog Low Isolation
............................................ 20 V normal operation, 400 V peak transient
EMC
........................................................................ Complies with EN/IEC 61326-1:2006. If used in areas with
Electromagnetic fields of 1 to 3 V/m, resistance outputs have a floor
adder of 0.508
Ω
. Performance not specified above 3 V/m. This
instrument may be susceptible to electro-static discharge (ESD) from
direct contact to the binding posts. Good static aware practices should
be followed when handling this and other pieces of electronic
equipment.
Line Power
............................................................. Line Voltage (selectable): 100 V, 120 V, 220 V, 240 V
Line Frequency: 47 Hz to 63 Hz
Line Voltage Variation:
±
10 % about line voltage setting
For optimal performance at full dual outputs (e.g. 1000 V, 20 A) choose
a ling voltage setting that is
±
7.5 % from nominal.
Power Consumption
............................................. 600 VA
Dimensions (HxWxL)
............................................ 17.8 cm x 43.2 cm x 47.3 cm (7 in x 17 in x 18,6 in) Standard rack
width and rack increment, plus 1.5 cm (0.6 in) for feet on bottom of
unit.
Weight (without options)
...................................... 22 kg (49 lb)
Absolute Uncertainty Definition
.......................... The 5522A specifications include stability, temperature coefficient,
linearity, line and load regulation, and the traceability of the external
standards used for calibration. You do not need to add anything to
determine the total specification of the 5522A for the temperature
range indicated.
Specification Confidence Interval
....................... 99 %
Detailed Specifications
DC Voltage
Absolute Uncertainty, tcal
±
5
°
C
±
(ppm of
μ
V)
Stability
Range
90 days
1 year
24 hours,
±
1
°
C
±
(ppm,
μ
V)
Resolution
μ
V
Max Burden
[1]
0 to 329.9999 mV
15 + 1
20 + 1
3 + 1
0.1
65
Ω
0 to 3.299999 V
9 + 2
11 + 2
2 + 1.5
1
10 mA
0 to 32.99999 V
10 + 20
12 + 20
2 + 15
10
10 mA
30 to 329.9999 V
15 + 150
18 + 150
2.5 + 100
100
5 mA
100 to 1020.000 V
15 + 1500
18 + 1500
3 + 300
1000
5 mA
Auxiliary Output (dual output mode only)
[2]
0 to 329.9999 mV
300 + 350
400 + 350
30 + 100
1
5 mA
0.33 to 3.299999 V
300 + 350
400 + 350
30 + 100
10
5 mA
3.3 to 7 V
300 + 350
400 + 350
30 + 100
100
5 mA
TC Simulate and Measure in Linear 10
μ
V/
°
C and 1 mV/
°
C modes
[3]
0 to 329.9999 mV
40 + 3
50 + 3
5 + 2
0.1
10
Ω
[1]
Remote sensing is not provided. Output resistance is <5 m
Ω
for outputs
≥
0.33 V. The AUX output has an output resistance of
<1
Ω
. TC simulation has an output impedance of 10
Ω
±
1
Ω
.
[2]
Two channels of dc voltage output are provided.
[3]
TC simulating and measuring are not specified for operation in electromagnetic fields above 0.4 v/m.