5522A
Getting Started
8
Connecting to Line Power
XW
Warning
To prevent possible electrical shock, fire, or personal injury:
•
Connect an approved three-conductor mains power cord to
a grounded power outlet.
•
Make sure that the Product is grounded before use.
•
Do not use an extension cord or adapter plug.
The calibrator is shipped with the appropriate line power plug for the country of
purchase. If you need a different type, refer to Table 3 and Figure 4 for a list and
illustration of the line power plug types available from Fluke.
After you verify that the line voltage selection is set correctly and that the correct fuse for
that line voltage is installed, connect the calibrator to a properly grounded three-prong
outlet.
How to Select Line Frequency
The calibrator is shipped from the factory for nominal operation at 60 Hz line frequency.
If you are using 50 Hz line voltage, you should re-configure the 5522A for optimal
performance at 50 Hz. To do so, from the front panel, go into SETUP, INSTMT SETUP,
OTHER SETUP, and then push the softkey under MAINS to change to 50 Hz. Store the
change. After the instrument is properly warmed up (on for 30 minutes or longer), you
must re-zero the complete instrument. For details, see the section on “Zeroing the
Calibrator” in Chapter 4.