5322A
Calibration Manual
78
Table 22. Two-wire Low Resistance Source
Nominal value
Lower limit
[1]
Upper limit
[1]
17
Ω
Rcal -45
m
Ω
Rcal +45
m
Ω
47
Ω
Rcal -300
m
Ω
Rcal +300
m
Ω
90
Ω
Rcal -500
m
Ω
Rcal +500
m
Ω
170
Ω
Rcal -1
Ω
Rcal
+1
Ω
470
Ω
Rcal -2.5
Ω
Rcal +2.5
Ω
900
Ω
Rcal -5
Ω
Rcal +5
Ω
1700
Ω
Rcal -10
Ω
Rcal +10
Ω
[1] Rcal is a calibration value of a selected standard.
Nominal value
Lower limit
Upper limit
100 m
Ω
.49 m
Ω
0.15 m
Ω
200 m
Ω
149.4 m
Ω
250.6 m
Ω
400 m
Ω
348.8 m
Ω
451.2 m
Ω
800 m
Ω
747.6 m
Ω
852.4 m
Ω
1
Ω
0.947
Ω
1.053
Ω
2
Ω
1.944
Ω
2.056
Ω
4
Ω
3.938
Ω
4.062
Ω
8
Ω
7.926
Ω
8.074
Ω
10
Ω
9.92
Ω
10.08
Ω
20
Ω
19.89
Ω
20.11
Ω
40
Ω
39.87
Ω
40.13
Ω
80
Ω
79.79
Ω
80.21
Ω
100
Ω
99.75
Ω
100.25
Ω
200
Ω
199.6
Ω
200.4
Ω
400
Ω
399.2
Ω
400.8
Ω
800
Ω
798.4
Ω
801.6
Ω
1 k
Ω
998
Ω
1002
Ω
2 k
Ω
1996
Ω
2004
Ω
4 k
Ω
3992
Ω
4008
Ω
8 k
Ω
7984
Ω
8016
Ω
10 k
Ω
9980
Ω
10020
Ω
Table 21. Four-wire Ground Bond Decade and Loop/Line Decade Test (cont.)