Voltmeter Adjustment HVR V01-HVR V04
Adjustment of the voltage auxiliary voltmeter requires that you set four constants in one
voltage range of 1000 V dc.
To make this adjustment:
1. Connect the 5522A or similar to the OUTPUT HI
Ω
HI and LO terminals, see
.
Figure 29. High Resistance Source Meters 1.5 kV Adjustment Connections 1
2. Select
High Resistance Meters
from calibration menu and push the
Select
softkey.
3. Select
HVR V01 Offset 1000Vdc (>1M)
and push the
Select
softkey.
4. Set the output voltage at 0.0 V dc on the 5322A and switch the output terminals ON.
5. Use the rotary knob or cursor keys to set the Product to 0.0 V.
6. Push the
Write
softkey.
7. Select
HVR V02 Voltage 1000Vdc (>1M)
and push the
Select
softkey.
8. On the 5522A, set the output voltage to 1000 V dc and switch the output terminals ON.
9. Use the rotary knob or the cursor keys to set the Product to1000.0 V.
10. Push the
Write
softkey.
11. Repeat this procedure similarly for the next two adjustment procedure positions (
HVR V03
Offset 1000Vdc (<1M)
and
HVR V04 Voltage 1000Vdc (<1M)
).
1
4
7
/
+
STBY
5522A CALIBRATOR
Fluke 5322A Fluke 5522A