Demo Instructions
5
Warning: defibrillators/external pacemakers are capable of emitting lethal
high voltages/energies and should be operated only by trained professionals.
DEFIBRILLATION TESTS
THE ENERGY TEST
1. Select DEFIB mode on the left panel, then select a test.
a.
Press
F1 ENERGY
b.
Press
F1
Set Wave and then select
VFib
” from list using the UP/DOWN keys
c.
Press F2 Wave Conversion and select “convert” from list using the UP/DOWN
keys
2. Place the defibrillator paddles securely on the analyzer contact plates to discharge the
defibrillator or use Optional Accessories (disposable pad adapters) by removing the two
defib paddle plug-in plates.