Demo Instructions

 

 

 

Warning: defibrillators/external pacemakers are capable of emitting lethal 

high voltages/energies and should be operated only by trained professionals. 

 
 

DEFIBRILLATION TESTS 
 
THE ENERGY TEST  

 
1. Select DEFIB mode on the left panel, then select a test.   
 

 

 

a.

 

Press 

F1 ENERGY

 

b.

 

Press 

F1

 Set Wave and then select 

VFib

” from list using the UP/DOWN keys 

 

 

 

c.

 

Press F2 Wave Conversion and select “convert” from list using the UP/DOWN 
keys   

 
2. Place the defibrillator paddles securely on the analyzer contact plates to discharge the 
defibrillator or use Optional Accessories (disposable pad adapters) by removing the two 
defib paddle plug-in plates.  

Summary of Contents for IMPULSE 6000D

Page 1: ...eart rate setting facilitates rate meter alarm testing 1 BPM per step 7 Easy to use connection posts for 12 Lead ECG 10 electrodes 8 Wide range of Arrhythmia selections 9 Auto range defib pacer measur...

Page 2: ...d professionals 3 Future upgradeable firmware software and new waveforms 4 Global support network Low level ECG Power switch Navigator Function Pacemaker input 50 to 1500 ohms t LCD 1 Plug in defib pa...

Page 3: ...hbutton The main menu appears on the analyzer s display The battery ICON in the upper right of the display indicates the relative charge level of the analyzer s internal battery If ICON is showing les...

Page 4: ...capable of emitting lethal high voltages energies and should be operated only by trained professionals 3 To set up unit for test apply ECG leads to the ten AHA color on the left side of the post and...

Page 5: ...B mode on the left panel then select a test a Press F1 ENERGY b Press F1 Set Wave and then select VFib from list using the UP DOWN keys c Press F2 Wave Conversion and select convert from list using th...

Page 6: ...onals 3 The analyzer calculates the energy delivered over the load resistance and displays the result in joules under the result Press F4 for summary 4 Show on the monitor that ECG waveform is convert...

Page 7: ...are available in the analyzer for performing this test Atrial fibrillation AFIB is the most representative clinical waveform for demonstration purposes Two selections are available AFIB Coarse is the...

Page 8: ...be operated only by trained professionals 5 Power on the defibrillator Select the desired energy place the paddles and discharge Note for the purpose of this demonstration AFIB is converted to a norma...

Page 9: ...ty in the defibrillator 1 Select F3 for CHARGE TIME test 2 Select F3 for MEASURE a Notice CHARGE TIME screen counting down 5 4 3 2 1 3 Select desired energy level and then charge defibrillator rillato...

Page 10: ...d and sensitivity tests measures and displays refractory periods etc 1 Connect pacer to the analyzer through either defibrillator input 50 ohms only or pacer input 50 to 1500 ohms 2 Select PACER mode...

Page 11: ...Turn pacer on and adjust pacer rate and output 8 Observe pacer Measurement result screen 9 Select F4 to view more tests F1 SENSITIVITY F2 REFRACTORY PERIOD Sensitivity test Demand sensitivity is the...

Page 12: ...MANCE can choose different wave forms such as square pulse sine and triangle c F3 PACER INTERACTIVE Selections Under F3 are a Non capture the analyzer operates in continuous mode However every 10th pa...

Page 13: ...e operated only by trained professionals b F2 PREMATURE c F3 VENTRICULAR d F4 MORE b Selects F4 again a Wave types i F1 CONDUCTION ii F2 TV PACED iii F3 R WAVE DETECTION iv Noise Immunity For more inf...