HF040TEN_v0402_03 Atex_IECEx_CSA_FM
Page 26
Data protection
Type
EEPROM backup of all settings. Data retention at least 10 years.
Pass code
Configuration settings can be pass code protected.
Hazardous area
(option)
Intrinsically safe
Type XI
ATEX approval:
II 1 G Ex ia IIC T4
II 1 D Ex iaD 20 IP 65 / 67 T 100°C
IECEx approval:
Ga Ex ia IIC T4
Ex iaD 20 IP 65 / 67 T 100°C
CSA / FM approval :
IS Class I/II/III, Division 1 Groups A to G T4
Class I zone 0 AEx ia IIC T4
Explosion proof
Type XF
ATEX approval ref.: <EX> II 2 GD EEx d IIB T5. Weight appr. 15kg.
Dimensions of enclosure: 350 x 250 x 200mm (13.7” x 9.9” x 7.9”) LxHxD.
Environment
Electromagnetic
compatibility
Compliant ref: EN 61326 (1997), EN 61010-1 (1993)
Low voltage directive
Compliant ref: EN60950.
Sensor
Type T
PT100 - 2, 3 or 4 wire
Accuracy Resolution: 16 bit. Error < 0.5°C
Offset -999.9 to +999.9 units
Update time Once per second.
Linearity Internally compensated
Operator functions
Displayed functions
•
temperature.
•
measuring unit
Temperature
Digits
5 ½ digits.
Units
°C - °F - K.
Decimals
1
INPUTS
OPERATIONAL