INSTALLATION
ST51A/ST75A/ST75AV Mass Flow Meters
18
Fluid Components International LLC
Figure 16 – Recommended Wire Routing/Internal Ground Screw
Accessing the Interface Board Connection Terminals
Warning:
Turn OFF instrument power source before wiring the instrument.
Caution:
Use caution inserting wires into electronics housing. The metal ends can damage circuit boards.
Remote Units: Avoid pulling, or inadvertently tugging, the remote cable when wiring the instrument. The sensor
connector/circuit board can be easily damaged by excess pulling of the remote cable.
To access the instrument’s connection terminals first use a .050
″
hex key to loosen the set screw locking the enclosure body blind lid (see
). Then unscrew the blind lid from the enclosure. Carefully pull the power and signal wires through the port to avoid
damaging the electronics.
Connect wiring as shown in the diagram in
and the summary list in
. Reinstall the blind lid when done
making the connections: Tighten the lid one full turn past the point where the O-ring makes contact with the lid, and then tighten the lid set
screw to lock the lid (set screw must not protrude from its threaded hole after tightening).
ESD Precautions
Caution:
FCI flow meters contain static-sensitive devices. To avoid damage to the instrument observe the ESD precautions
listed below before opening the instrument for wiring.
●
Use a wrist band or heel strap with a 1
MΩ
resistor connected to ground.
●
Use a static conductive mat on the work table or floor with a 1
MΩ
resistor connected to the ground when working on the instrument in
a shop setting.
●
Connect the instrument to ground.
●
Apply antistatic agents such as Static Free made by Chemtronics to hand tools used on the instrument.
●
Keep high static-producing items away from the instrument.
C01162-1-2
INTERNAL GROUND
SCREW, 10-32 x ¼
″
SLOTTED HEX WASHER
POWER WIRING
SIGNAL WIRING
ENCLOSURE BODY
(BLIND LID REMOVED,
AC VERSION SHOWN)
Summary of Contents for ST51A
Page 1: ...Doc No 06EN003426 Rev C...
Page 83: ...ST51A ST75A ST75AV Mass Flow Meters APPENDIX A DRAWINGS Fluid Components International LLC 75...
Page 84: ...APPENDIX A DRAWINGS ST51A ST75A ST75AV Mass Flow Meters 76 Fluid Components International LLC...
Page 85: ...ST51A ST75A ST75AV Mass Flow Meters APPENDIX A DRAWINGS Fluid Components International LLC 77...
Page 86: ...APPENDIX A DRAWINGS ST51A ST75A ST75AV Mass Flow Meters 78 Fluid Components International LLC...
Page 87: ...ST51A ST75A ST75AV Mass Flow Meters APPENDIX A DRAWINGS Fluid Components International LLC 79...
Page 88: ...APPENDIX A DRAWINGS ST51A ST75A ST75AV Mass Flow Meters 80 Fluid Components International LLC...
Page 89: ...ST51A ST75A ST75AV Mass Flow Meters APPENDIX A DRAWINGS Fluid Components International LLC 81...
Page 90: ...APPENDIX A DRAWINGS ST51A ST75A ST75AV Mass Flow Meters 82 Fluid Components International LLC...
Page 91: ...ST51A ST75A ST75AV Mass Flow Meters APPENDIX A DRAWINGS Fluid Components International LLC 83...
Page 92: ...APPENDIX A DRAWINGS ST51A ST75A ST75AV Mass Flow Meters 84 Fluid Components International LLC...
Page 93: ...ST51A ST75A ST75AV Mass Flow Meters APPENDIX A DRAWINGS Fluid Components International LLC 85...
Page 94: ...APPENDIX A DRAWINGS ST51A ST75A ST75AV Mass Flow Meters 86 Fluid Components International LLC...
Page 95: ...ST51A ST75A ST75AV Mass Flow Meters APPENDIX A DRAWINGS Fluid Components International LLC 87...
Page 96: ...APPENDIX A DRAWINGS ST51A ST75A ST75AV Mass Flow Meters 88 Fluid Components International LLC...
Page 97: ...ST51A ST75A ST75AV Mass Flow Meters APPENDIX A DRAWINGS Fluid Components International LLC 89...