![Fluid Components Intl ST51A Installation, Operation & Maintenance Instruction Manual Download Page 18](http://html1.mh-extra.com/html/fluid-components-intl/st51a/st51a_installation-operation-and-maintenance-instruction-manual_2308788018.webp)
INSTALLATION
ST51A/ST75A/ST75AV Mass Flow Meters
10
Fluid Components International LLC
Retraction/Removal Procedure
1. Loosen the socket head cap screw on the side of the locking collar. See
Caution:
For applications where the process media is pressurized to greater than 232 psig [16 bar(g)] make sure to first
depressurize the process line before retracting the flow element. At 232 psig [16 bar(g)], the effective force on the
insertion probe is 45.5 lbs (20.6 kg), which is the limit at which the flow element can be safely guided by hand. When
using hands to restrain the retraction, be prepared for a rapid pressure impulse of the flow element. Make sure that
there are no objects directly behind the flow element as the insertion probe may retract very quickly.
Figure 7 – Retractable Packing Gland Locking Collar Detail
2. Slowly loosen the packing nut until the insertion probe begins to retract. Use hands as needed to help control the retraction. If the
probe does not begin to retract itself, gently shake and pull the insertion probe until the flow element has been fully retracted into the
packing gland.
3. For ball valve applications, close the ball valve immediately after retraction to seal off the process. After closing the ball valve it is then
safe to remove the flow element from the back end of the ball valve. If a ball valve is not being used, make sure to first
depressurize the process line before removing the flow element.
PACKING NUT
LOOSEN THIS SCREW
LOCKING COLLAR
C01445-1-1
Summary of Contents for ST51A
Page 1: ...Doc No 06EN003426 Rev C...
Page 83: ...ST51A ST75A ST75AV Mass Flow Meters APPENDIX A DRAWINGS Fluid Components International LLC 75...
Page 84: ...APPENDIX A DRAWINGS ST51A ST75A ST75AV Mass Flow Meters 76 Fluid Components International LLC...
Page 85: ...ST51A ST75A ST75AV Mass Flow Meters APPENDIX A DRAWINGS Fluid Components International LLC 77...
Page 86: ...APPENDIX A DRAWINGS ST51A ST75A ST75AV Mass Flow Meters 78 Fluid Components International LLC...
Page 87: ...ST51A ST75A ST75AV Mass Flow Meters APPENDIX A DRAWINGS Fluid Components International LLC 79...
Page 88: ...APPENDIX A DRAWINGS ST51A ST75A ST75AV Mass Flow Meters 80 Fluid Components International LLC...
Page 89: ...ST51A ST75A ST75AV Mass Flow Meters APPENDIX A DRAWINGS Fluid Components International LLC 81...
Page 90: ...APPENDIX A DRAWINGS ST51A ST75A ST75AV Mass Flow Meters 82 Fluid Components International LLC...
Page 91: ...ST51A ST75A ST75AV Mass Flow Meters APPENDIX A DRAWINGS Fluid Components International LLC 83...
Page 92: ...APPENDIX A DRAWINGS ST51A ST75A ST75AV Mass Flow Meters 84 Fluid Components International LLC...
Page 93: ...ST51A ST75A ST75AV Mass Flow Meters APPENDIX A DRAWINGS Fluid Components International LLC 85...
Page 94: ...APPENDIX A DRAWINGS ST51A ST75A ST75AV Mass Flow Meters 86 Fluid Components International LLC...
Page 95: ...ST51A ST75A ST75AV Mass Flow Meters APPENDIX A DRAWINGS Fluid Components International LLC 87...
Page 96: ...APPENDIX A DRAWINGS ST51A ST75A ST75AV Mass Flow Meters 88 Fluid Components International LLC...
Page 97: ...ST51A ST75A ST75AV Mass Flow Meters APPENDIX A DRAWINGS Fluid Components International LLC 89...