Flow Controller Systems FCS KA SS Operation Manual Download Page 5

   

 

   

 

KA (SS)

 

 

 

- 5 - 

(2)

 

In the stop state, press the digital knob, the lower right corner of the OLED flashes and the value is RS485 communication address 

(1~32). At this time, you can turn the digital knob to change the value. After setting, press the digital knob to save the data and the 

screen no longer flashes. 

(3)

 

Press the digital knob and in same time turn on the pump to check the program version number and the SN code of the pump. 

4.

 

  External Control Instruction 

 

.

 

RS232 Communication Port 

Serial 232 communication interface, this interface is directly connected to the computer’s serial interface, the protocol is Modbus 

protocol. 

.

 

Analog Signal Input Port

: Control the motor speed from 0 rpm to maximum speed through analog signal. 

0-5V/0-10V

: 0V to 5V/10V voltage signal input port. 

First, short-circuited the external control identification port (IN3) and GD2, the default is 0-5V speed control. Connect the 0-5V pin of the 

green terminal to the positive pole of the analog signal, and connect the GND pin to the negative pole of the analog signal. At this time, 

adjusting the analog signal speed will change accordingly. 

4-20mA

: 4 to 20mA current signal input port 

First, short-circuited the external control identification port (IN3) and the 0-5V/4-20mA speed control conversion interface (IN4) with GD2, 

at this time it is 4-20mA speed control. Connect the 4-20mA pin of the green terminal to the positive pole of the analog signal, and connect 

the GND pin to the negative pole of the analog signal. At this time, adjusting the analog signal speed will change accordingly. 

GND

: Analog signal negative terminal. 

Note

:   

Ø

 

The factory default of voltage analog speed control is 0-5V. If customers need 0-10V speed control, please contact us for technical 

support. 

Ø

 

It is forbidden to connect the 0-5V signal to the 4-20mA input terminal, the wrong connection may cause damage to the equipment. 

The external control port plug must be plugged and unplugged when the power is off to prevent the external control port from being 

burned. 

.

 

RS485 Communication Port 

B-

 

: Connect to RS485 B- 

GND

TXD

RXD

VDD2

0 - 5 V

4 - 20

mA

GD1

启停

B -

A +

RS232

接口

RS485

接口

GD2

启停

状态输出

外控识别端口

0-5V/4-20mA

调速转换接口

POWER_VDD

POWER_COM

GD2

+24V

GND

IN1

IN2

IN3

IN4

RL1

1

2

3

4

5

换向——对GD2换向

Status 

Output

0-5V/4-20mA Speed Conversion Interface

External Control Identification Port

Change Direction

Change Direction for GD2

Start/Stop

For GD2

RS485 Interface

RS232 Interface

Summary of Contents for FCS KA SS

Page 1: ...FCS KA SS Compact Peristaltic Pump Operation Manual 0 1 2 3456 1 7 7 8 9 88 1 6 A B6 1 C46DEF GH C46DEF GH ...

Page 2: ...e Ø If the power cord or plug had wear and other damage please disconnect the plug Hold the plug instead of the wire Ø If following situations happened please turn off the power supply and disconnect the plug Hold the plug instead of the wire 1 Fluid splash on the pump 2 You think the pump need to maintain or repair Ø The user s power socket must have ground wire and a reliable grounding Note The ...

Page 3: ... Introduction 4 2 Product Appearance 4 3 Operation Panel 4 4 External Control Instruction 5 5 Technical Specification 7 6 Main Function and Feature 8 7 Dimension Drawing 8 8 Maintenance 9 9 Warranty After Sales Service 9 ...

Page 4: ...ntrol the pumps start stop press this button once the pump working status change once When drive runs the screen display when the pump stops the screen display Press the Start stop button and in same time turn on the pump to reset the pump Ø Change Direction Button Control the pump running direction press the change direction button once the pump working direction change once Ø Digital Knob The di...

Page 5: ...gly 4 20mA 4 to 20mA current signal input port First short circuited the external control identification port IN3 and the 0 5V 4 20mA speed control conversion interface IN4 with GD2 at this time it is 4 20mA speed control Connect the 4 20mA pin of the green terminal to the positive pole of the analog signal and connect the GND pin to the negative pole of the analog signal At this time adjusting th...

Page 6: ...IN3 External control identification port after the external control isolation power supply is added between VDD2 and GD2 the IN3 pin and GD2 pin are short circuited for the external control speed adjustment identification and there are related indications on the OLED screen IN4 External control speed 0 5V 4 20mA switch Add an external control isolation power supply between VDD2 and GD2 and short c...

Page 7: ...eed control 0 5V 0 10V 4 20mA for option Speed resolution 0 1rpm Communication RS485 RS232 Control Digital knob and keypad Temperature 0 40 Motor Stepper motor Relative humidity 80 Display OLED Output interface Output motor working status Consumption 15W IP rate IP31 GND TXD RXD VDD2 GD1 B A POWER_VDD POWER_COM GD2 24V GND IN1 IN2 IN3 IN4 RL1 1 2 3 4 5 4 20mA 0 5V 5V电源 电流模 拟量信 号正极 电流模 拟量信 号负极 5V P...

Page 8: ...llent performance good heat dissipation conditions low working noise stable operation and has power off memory function Ø External control adopts standard terminal blocks which is more suitable for supporting industrial use Ø RS485 standard Modbus protocol more convenient to connect with PLC Ø 304 stainless steel shell anti corrosion no rust conform to GMP requirements Ø OLED display motor speed a...

Page 9: ... indirectly related to or arising out of the use of its products This warranty does not obligate our company to bear any costs of removal installation transportation or other charges which may arise in connection with a warranty claim If the pump fails during the warranty period after confirmation by our technical department we will provide spare parts free of charge Customers will need to bear th...

Reviews: