Field Firmware Updates
The USB port in the battery compartment allows for field firmware updates,
follow the steps below.
1. Connect the CM65 to a PC using a USB cable. The CM65 USB port is lo-
cated in the battery compartment.
2. Turn the CM65 rotary switch to any position, the CM65 will display ‘USB’.
3. Download the update file (CM65_V0.0x.hex) from the FLIR support site
(https:\\support.flir.com) and transfer it to the CM65 internal memory root
directory.
4. IMPORTANT: Disconnect the CM65 from the PC. Do not proceed unless
the USB cable that connects the CM65 to the PC is disconnected.
5. Turn the CM65 rotary switch to the OFF position.
6. Turn the CM65 rotary switch to any ON position.
7. The CM65 will begin the update process. This will take approximately 30
seconds during which there will be a blank screen.
8. When the update is complete, the CM65 will emit an alert beep and will
reboot.
9. This completes the update process, if errors occur please contact FLIR
customer support.