A Circuit Test is to be performed each time a Patient Circuit or components
of a patient Circuit are replaced:
2
1.
Connect an adult (500 ml) test lung with a 90 degree
elbow to the Flow Sensor’s patient side.
2.
Press the
Technical
button once, and then
tap the Circuit TEST button.
3.
Press
Enter
Circuit TEST