14 User modes
PIOX S72*
137
UMPIOX_S72xV1-9EN, 2022-05-15
14
User modes
The user modes allow extended diagnostics of signals and
measured values as well as the definition of additional
parameters adapted to the application.
The following user modes can be selected:
•StandardUser
•ExpertUser
•SuperUser
•SuperUser ext.
Depending on the selected user mode, further options are displayed in the program branch
Options\Channel\
Special settings
.
Special functions
System settings
User mode
→
StandardUser
→
ExpertUser
→
SuperUser
→
SuperUser ext.
see annex A
Tab. 14.1: Menu items of the user modes
Options\Special settings
StandardUser
ExpertUser
SuperUser
SuperUser ext.
default
Cut-off flow
x
x
x
x
On
Flow velocity limit
x
x
x
Off
Pig detection
x
x
Off
Turbulence mode
x
x
x
x
Off
Max. amplification
x
x
x
Off
Pipe signal detection
x
x
x
On
LWT pipe wall calibr.
x
x
Off
Linear calibration
x
x
Off
Profile correction
x
x
x
On
Weighting factor
x
x
Off
Multi-point calibration
(if enabled in
Special
functions\Measurement\
Measurement settings
)
x
x
x
x
Transducer temp.
and
Transd.
temp. violat.
(as source item
Diagnostic values
)
x
x
Extended diagnostics
(in the program branch
Options
,
calculation channels)
x
x
x