8.2
Power supply
The external protective earth is connected to the equipotential bonding terminal on the housing of the transmitter FLUXUS
F704**-A2 (see Fig. 8.13). It must always be connected.
For the connection of the power cable to the transmitter, see box "Cable connection", Fig. 8.13 and Tab. 8.10.
For the voltage, see Technical specification.
Attention!
Observe the "Safety instructions for the use in explosive atmospheres" (see SIFLUXUS document).
Attention!
According to IEC 61010-1:2010, a switch has to be provided near the measuring instrument in the
building installation, easily accessible for the user and marked as a disconnection device for the mea-
suring instrument.
If the transmitter FLUXUS F704**-A2 is used in an explosive atmosphere, the switch should be in-
stalled outside the explosive atmosphere. If this is not possible, the switch should be installed in the
least hazardous area.
Attention!
The degree of protection of the transmitter will only be guaranteed if the power cable fits firmly and
tightly in the cable gland.
Cable connection
• Remove the blind plug from the transmitter to connect the cable (if present).
• Prepare the cable with an M20 cable gland. The used cable has to have a wire cross-section of 0.5…1.5 mm².
• Push the cable through the cap nut, compression part and basic part of the cable gland.
• Insert the cable into the housing of the transmitter.
• Screw the sealing ring side of the basic part tightly into the housing of the transmitter.
• Fix the cable gland by screwing the cap nut onto the basic part.
• Connect the cable to the terminals of the transmitter.
Fig. 8.13: Transmitter FLUXUS WW
Tab. 8.10: Terminal assignment
terminal
connection AC
connection DC
PE
earth
earth
N(-)
neutral
-DC
L(+)
phase 100...230 V AC, 50...60 Hz
+DC
fuse
1 A, time-lag
1.6 A, time-lag
AVS
AV
AGN ARS AR
BVS
BV
BGN BRS BR
SA1 SA2 SA3 SA4
SB1 SB2 SB3 SB4
T1a T1b S2 T2a
T3a T3b S4 T4a
T2b
T4b
T1A T1B S1 T2A
T3A T3B S3 T4A
T2B
T4B
A+ B- P1+ P2+
P4+ P5a P6a P7a
P3+
101 103 P1- P2-
P4- P5b P6b P7b
P3-
PE N(-) L(+)
cable gland
cap nut
compression part basic part
power supply
equipotential bonding terminal
(FLUXUS F704**-A2)
Summary of Contents for FLUXUS WW
Page 1: ...User manual UMFLUXUS_WWV1 1EN Ultrasonic flowmeter for water and wastewater FLUXUS WW...
Page 4: ...FLUXUS WW UMFLUXUS_WWV1 1EN 2018 02 28...
Page 8: ...FLUXUS WW Table of contents 8 UMFLUXUS_WWV1 1EN 2018 02 28...
Page 144: ...FLUXUS WW D Conformity declarations 144 UMFLUXUS_WWV1 1EN 2018 02 28 D Conformity declarations...
Page 145: ......
Page 146: ......