14 Settings
101
14.2.2
Fluid pressure
The dependence of the parameters of a fluid on the pressure can be taken into account.
14.2.3
Measuring point number
14.2.4
Transducer distance
14.2.5
Error value delay
The error value delay is the time after which an error value will be sent to an output if no valid measured values are avail-
able.
14.2.6
Alarm state indication
During the next scroll through the program branch
Parameter
, the outer pipe diameter
that corresponds to the entered pipe circumference will be displayed.
Example: 180 mm :
π
= 57.3 mm
Note!
The edition of the pipe circumference is only temporarily. When the transmitter switches back to the
display of the pipe circumference (internal recalculation), slight rounding errors may occur.
Example:
entered pipe circumference: 100 mm
displayed outer pipe diameter: 31.8 mm
When the transmitter switches back to the display of the pipe circumference, 99.9 mm will be dis-
played.
If
on
is selected, the fluid pressure will be requested in the program branch
Parameter
.
If
off
is selected, 1 bar will be used for all calculations.
Note!
For documentation purposes, it is useful to enter the fluid pressure, even if the transmitter contains
no pressure-dependent characteristic curves.
Select (
1234
) if the measuring point is to be identified only by numbers, point and hyphen.
Select (
↑↓←
→
) if the measuring point is to be designated with ASCII characters.
recommended setting:
user
•
user
will be selected if the measuring point is always the same
•
auto
can be selected if the measuring point changes often
In the program branch
Measuring
, the recommended transducer distance will be dis-
played in parentheses, followed by the entered transducer distance if the recommended
and the entered transducer distance are not identical.
During the transducer positioning, in the program branch
Measuring
• only the entered transducer distance will be displayed if
Transd. Distance
=
user
is
selected and the recommended and the entered transducer distances are identical
• only the recommended transducer distance will be displayed if
Transd. Distance
=
auto
is selected
Select
edit
to enter an error value delay. Select
damping
if the damping factor is to be
used as the error value delay.
For further information on the behavior in case of missing measured values, see section
16.1.2 and 16.2.
Select
on
to display the alarm state during the measurement.
For further information on the alarm outputs, see section 16.5.
Outer Diameter
57.3
mm
Fluid pressure
off
>ON<
Meas.Point No.:
(1234) >(
↑↓←
→
)<
Transd. Distance
auto
>USER<
Transd. Distance?
(50.8) 50.0 mm
Transd. Distance?
50.8
mm
Error-val. delay
damping
>EDIT<
SHOW RELAIS STAT
off
>ON<
Summary of Contents for FLUXUS WW
Page 1: ...User manual UMFLUXUS_WWV1 1EN Ultrasonic flowmeter for water and wastewater FLUXUS WW...
Page 4: ...FLUXUS WW UMFLUXUS_WWV1 1EN 2018 02 28...
Page 8: ...FLUXUS WW Table of contents 8 UMFLUXUS_WWV1 1EN 2018 02 28...
Page 144: ...FLUXUS WW D Conformity declarations 144 UMFLUXUS_WWV1 1EN 2018 02 28 D Conformity declarations...
Page 145: ......
Page 146: ......