![Flexim FLUXUS WD100 User Manual Download Page 24](http://html1.mh-extra.com/html/flexim/fluxus-wd100/fluxus-wd100_user-manual_2304737024.webp)
6 Installation of the Transmitter
24
UMFLUXUS_F5WDV1-1EN, 2016-04-08
6.2.3
Pipe Installation
Installation on a 2 " pipe
• Fix the pipe mounting plate (2) to the pipe (see Fig. 6.3).
• Fix the instrument mounting plate (3) to the pipe mounting plate (2) with the nuts (4).
• Fix the bottom side of the housing to the instrument mounting plate (3).
Installation on a pipe > 2 "
The mounting kit is fixed to the pipe with tension straps (5) instead of the shackle (see Fig. 6.3.).
• Push the tension straps (5) through the holes in the instrument mounting plate (3).
Fig. 6.2: FLUXUS WD100, WD200 (dimensions in mm)
Fig. 6.3: Pipe installation set
Ø 8.4
165
10
10
3
Ø 4.5
screw holes
for wall mounting
1 shackle
2 pipe mounting plate
3 instrument mounting plate
4 nut
5 tension strap
1
2
3
4
5