7 Connection
FLUXUS G809
7.1 Transducers
63
UMFLUXUS_G809V2-2EN, 2020-06-25
7.1
Transducers
The transmitter has 2 cable glands for the connection of the transducers. If the transmitter has only one measuring chan
-
nel, one of the openings is closed with a blind plug.
It is recommended to run the cables from the measuring point to the transmitter before connecting the transducers to
avoid load on the connectors.
Transmitter FLUXUS G809**-A1
7.1.1
Connection of the transducer cable to the transmitter
7.1.1.1
Transducer cable with plastic cable jacket and stripped cable ends
• Remove the blind plug for the connection of the transducer cable.
• Open the cable gland of the transducer cable. The compression part remains in the cap nut.
• Push the transducer cable through the cap nut and the compression part.
• Prepare the transducer cable.
• Shorten the external shield and brush it back over the compression part.
• Screw the sealing ring side of the basic part into the transmitter housing.
• Insert the transducer cable into the housing.
Notice!
If transducers are replaced or added, the sensor module also has to be replaced or added, see section 7.1.8.
Fig. 7.2:
Connection of the transducers to the transmitter
1 – transducers (measuring channel A)
2 – transducers (measuring channel B)
Important!
The degree of protection of the transmitter is only ensured if all cables are tightly fitted using cable glands and the
housing is firmly screwed.