7 Connection
7.1 Transducers
FLUXUS G809
2020-06-25, UMFLUXUS_G809V2-2EN
74
7.1.8
Transducer module (SENSPROM)
The transducer module contains important transducer data for the operation of the transmitter with transducers.
If transducers are replaced or added, the sensor module must also be replaced or added.
• Disconnect the transmitter from the power supply.
• Insert the transducer modules into the sockets, see Fig. 7.13. Socket A is assigned to the transducers of measuring
channel A, socket B to the transducers of measuring channel B.
• Connect the transmitter to the power supply.
• Enter all parameters of the program branch
Parameter
.
• Start the measurement.
Notice!
The serial numbers of the transducer module and the transducer must be identical. A wrong or incorrectly connected
sensor module will lead to incorrect measured values or to a measurement failure.
Fig. 7.13: Connection of the transducer module
1 – transducer module (measuring channel A)
2 – transducer module (measuring channel B)