7 Connection
FLUXUS F808
7.1 Transducers
67
UMFLUXUS_F808V2-2EN, 2020-06-25
Transmitter FLUXUS F808**-F2
7.1.8
Connection of the transducer cable to the transmitter
For the installation of transducers with SMB connectors a SMB adapter is required.
• Remove the blind plug for the connection of the transducer cable.
• Connect the transducers with the SMB adapter.
• Insert the transducer in the SMB connector into the housing.
• Screw the sealing ring side of the basic part into the transmitter housing.
• Fix the cable gland by screwing the cap nut onto the basic part.
• Connect the cable of the SMB adapter to the terminals of the transmitter, see Fig. 7.13 and Tab. 7.12.
7.1.9
Connection of the extension cable to the junction box
• The extension cable is connected to the transmitter via the transducer connection.
• Remove the blind plug for the connection of the transducer cable.
• Open the cable gland of the extension cable. The compression part remains in the cap nut.
• Push the extension cable through the cap nut and the compression part.
• Prepare the extension cable.
• Shorten the external shield and brush it back over the compression part.
• Screw the sealing ring side of the basic part into the transmitter housing.
• Insert the extension cable into the housing.
• Fix the cable gland by screwing the cap nut onto the basic part.
• Connect the extension cable to the terminals of the transmitter, see Fig. 7.14 and Tab. 7.13.
Tab. 7.12: Terminal assignment
terminal
connection
AV
SMB adapter (brown cable, marked white)
AVS
SMB adapter (red cable)
ARS
SMB adapter (red cable)
AR
SMB adapter (brown cable)
Fig. 7.13: Connection of the transducer cable with SMB adapter to the transmitter
1 – transducers
2 – SMB adapter
Notice!
For good electromagnetic compatibility (EMC), it is important to ensure good electrical contact between the external
shield and the cap nut (and thus the housing).
1
ARS
AR
AV
AVS
L(+)
N(-)
PE
KL4
KL1
KL2
KL3
2