7 Connection
FLUXUS F808
7.1 Transducers
65
UMFLUXUS_F808V2-2EN, 2020-06-25
7.1.7
Connection to the junction box
Transducer and extension cables are connected via the terminal board KFM1. The terminal board has to be installed into
a junction box (by the customer) approved for the use in explosive atmospheres.
• Connect the transducer and extension cable to the terminals of the junction box, see Tab. 7.10, Tab. 7.11 and Fig. 7.12.
Fig. 7.11: Connection of the extension cable to the transmitter
1 – external shield brushed back
Tab. 7.10: Terminal assignment (transducer cable)
terminal
connection (KL1)
terminal
connection (KL2)
V
transducer (core)
R
transducer (core)
VS
transducer (internal shield)
RS
transducer (internal shield)
VOS
transducer (external shield)
ROS
transducer (external shield)
Tab. 7.11: Terminal assignment (extension cable)
terminal
connection (KL3)
TV
core
TVS
internal shield
TOS
external shield
TRS
internal shield
TR
core
70 mm
20 mm
10 mm
1
ARS
AR
AV
AVS
L(+)
N(-)
PE
KL4
KL1
KL2
KL3