7 Connection
7.1 Transducers
FLUXUS F808
2020-06-25, UMFLUXUS_F808V2-2EN
64
Messumformer FLUXUS F808**-F1
7.1.5
Connection of the transducer cable to the transmitter
• Remove the blind plug for the connection of the transducer cable.
• Insert the transducer cable through the cable conduit (approved for FM Class I, Division 1) into the housing.
• Connect the transducer cable to the terminals of the transmitter, see Fig. 7.10 and Tab. 7.8.
7.1.6
Connection of the extension cable to the transmitter
• The extension cable is connected to the transmitter via the transducer connection.
• Remove the blind plug for the connection of the transducer cable.
• Insert the extension cable through the cable conduit (approved for FM Class I, Division 1) into the housing.
• Prepare the extension cable.
• Connect the extension cable to the terminals of the transmitter, see Fig. 7.11 and Tab. 7.9.
Tab. 7.8:
Terminal assignment
terminal
connection
AV
transducer (core)
AVS
transducer (internal shield)
ARS
transducer (inner shield)
AR
transducer (core)
Fig. 7.10: Connection of the transducer cable to the transmitter
1 – transducer cable
Tab. 7.9:
Terminal assignment
terminal
connection
AV
white or marked cable (core)
AVS
white or marked cable (internal shield)
ARS
brown cable (internal shield)
AR
brown cable (core)
equipotential bonding terminal
external shield
1
ARS
AR
AV
AVS
L(+)
N(-)
PE
KL4
KL1
KL2
KL3