7 Connection
UMFLUXUS_F7V4-6-2EN, 2017-10-01
61
7.1.2.2
Extension Cable with Plastic Cable Jacket and Stripped Cable Ends
FLUXUS F704
• Remove the left blind plug for the connection of extension cable (see Fig. 7.14).
• Open the cable gland of the extension cable. The compression part remains in the cap nut.
• Push the extension cable through the cap nut and the compression part.
• Prepare the extension cable.
• Cut the external shield and brush it back over the compression part.
• Screw the gasket ring side of the basic part into the housing of the transmitter.
• Insert the extension cable into the housing.
• Fix the cable gland by screwing the cap nut onto the basic part.
• Connect the extension cable to the terminals of the transmitter (see Fig. 7.1, Fig. 7.14 and Tab. 7.6).
FLUXUS F705, F706
• Remove the left blind plug for the connection of extension cable (see Fig. 7.15).
• Open the cable gland of the extension cable. The compression part remains in the cap nut.
• Push the extension cable through the cap nut, the compression part, the basic part and the gasket ring.
(Sealing ring: only for cable gland M20, not for cable gland 1/2 NPS.)
• Prepare the extension cable.
• Cut the external shield and brush it back over the compression part.
• Insert the extension cable into the housing.
• Fix the cable gland by screwing the cap nut onto the basic part.
• Fix the extension cable by tightening the cable gland with the counter nut.
• Connect the extension cable to the terminals of the transmitter (see Fig. 7.2, Fig. 7.15 and Tab. 7.6).
Tab. 7.5: Terminal assignment
terminal
connection
measuring
channel A
measuring
channel B
X6AV
X8BV
SMB connector (white or marked cable)
X6AR
X8BR
SMB connector (brown cable)
X5
X7
AMP-Quick connector (connection of the sensor module)
Attention!
Observe the "Safety Instructions for the Use in Explosive Atmosphere" (see document SIFLUXUS).
Attention!
For good high frequency shielding, it is important to ensure good electrical contact between the exter-
nal shield and the cap nut (and the housing).
Attention!
For good high frequency shielding, it is important to ensure good electrical contact between the exter-
nal shield and the cap nut (and the housing).
Tab. 7.6: Terminal assignment
terminal
connection
AV
white or marked cable (core)
AVS
white or marked cable (shield)
ARS
brown cable (shield)
AR
brown cable (core)
Summary of Contents for FLUXUS F704**-A2 Series