PF7/PF7+ Peristaltic Filler with QC14 Pumphead User Manual
20.17.1 Setting fill tolerance limits
The fill tolerance limits specified in "Guideline fill tolerance limits" below are
calculated based on the expected smallest drip size. These are a minimum
recommended tolerance. You may set a smaller tolerance than those specified,
however doing so may result in more unacceptable fills, particularly during the initial
fills of a zero waste batch. Setting a smaller tolerance limit will not increase the
accuracy of the fill, it will only determine the limits that define an acceptable fill.
To define the fill tolerances in ml:
1.
In the recipe menu select Fill tolerance.
2.
Select Upper limit or Lower limit as required
3.
%
4.
Enter the value specified in the table as a percentage of the fill volume defined
in the recipe
Table 12 - Guideline fill tolerance limits
Tube size (mm)
Recommended tolerance in (ml)
0.5
±0.0099
0.8
±0.0149
1.2
±0.0149
1.6
±0.0209
3.2
±0.0407
4.8
±0.0594
6
±0.0792
8
±0.1100