WARNINGS:
•
Portable RF communications equipment (including peripherals such as antenna cables and
external antennas) should be used no closer than 30 cm (12 inches) to any part of the M1 System,
including cables specified by the manufacturer. Otherwise, degradation of the performance of
this equipment could result.
•
The M1 should not be used adjacent to or stacked with other equipment. If adjacent or stacked
use is necessary, the M1 should be observed to verify normal operation. If operation is not normal,
the M1 or the other equipment should be moved.
•
Use of accessories, transducers and cables other than those specified or provided by the
manufacturer of this equipment could result in increased electromagnetic emissions or decreased
electromagnetic immunity of this equipment and result in improper operation.
Emissions
Electromagnetic Emissions
The M1 is intended for use in the electromagnetic environment specified below. The customer or the
user of the M1 should assure that it is used in such an environment.
Emission Tests
Compliance
Electromagnetic Environment
–
Guidance
RF emissions
CISPR 11
Group 1
The M1 uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely
to cause any interference in nearby electronic equipment.
RF emissions
CISPR 11
Class B
The M1 is suitable for use in all establishments, including
domestic establishments and those directly connected to
the public low-voltage power supply network that supplies
buildings used for domestic purposes.
Immunity
Electromagnetic Immunity
The M1 is intended for use in the electromagnetic environment specified below. The customer or the
user of the M1 should assure that it is used in such an environment.
Immunity Tests
Compliance Level
Electromagnetic Environment
–
Guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 8kV contact
± 15kV air
The relative humidity should be at least 5 %
Power frequency
(50/60 Hz)
magnetic field IEC
61000-4-8
30A/m
Power frequency magnetic fields from common appliances
in the home are not expected to affect the device.
Power frequency magnetic fields should be at levels
characteristic of a typical location in a typical commercial
or hospital environment. Keep the M1 away from sources
of high levels of power line magnetic fields (in excess of 30
A/m) to reduce the likelihood of interference.
NOTE: UT is the A/C. mains voltage prior to application of the test level.