DIRECTIONS
followed by last recorded temperature. After showing the self-test temperature, the thermometer is
now in the testing mode.
1. Press the
On/Off
Button next to LCD display. A tone will sound as the screen shows ,
*Note: If the measured temperature is less than 32 or 90 , the LCD will display Lo, if the measured
temperature is higher than 42.9 or 109.9 , the LCD will display Hi.
℃
℉
℃
℉
2. Position thermometer in desired location (mouth, rectum, or armpit.)
a)
O r al U se:
Place thermometer under tongue as indicated by
position shown in Figure 2. Close your mouth and breathe evenly
through the nose to prevent the measurement from being influenced
by inhaled/exhaled air. Normal temperature between 35.7 and 37.3
(96.3 and 99.1 )
b)
R ectal U se:
Lubricate silver probe tip with petroleum jelly for easy insertion. Gently insert
sensor approximately 1cm (less than 1/2 ) into rectum. Normal temperature between 36.2
and 37.7 (97.2 and 99.9 ).
c)
A r mpit U se:
Wipe armpit dry. Place probe in armpit and keep arm pressed firmly at side.
From a medical viewpoint, this method will always provide inaccurate readings, and should not
be used if precise measurements are required. Normal temperature between 35.2 and 36.7
(95.4 and 98.1 ).
3. The degree sign flashes throughout the testing process. When flashing stops an alarm will beep for
approximately 10 seconds. The measured reading will appear on the LCD simultaneously. The
minimum measurement time until the signaling tone (beep) must be maintained without exception. The
measurement continues even after the buzzer notification. The displayed temperature will not change
when thermometer is removed from its testing position.
4. To prolong battery life, press the
On/Off
Button to turn unit off after testing is complete. If no action
is taken, the unit will automatically shut off after around10 minutes.
5. Store the thermometer in its protective case.
“
√
”
℃
℃
℉
℉
"
℃
℃
℉
℉
℃
℃
℉
℉
*Note: Normally
Alarm b eeps m ore ra pidly w hen te mperature
reaches 37.8 (100 ) or higher ,and t
the buzzes are
“
Bi
-
Bi
-
Bi
-
B i
-
”
;
℃
℉
he buzzes are
“
Bi-Bi-Bi----- Bi-Bi-Bi------- Bi-Bi-Bi
”
BATTERY REPLACEMENT
1. Replace battery when
appears in the lower right corner of LCD display.
2. Pull battery cover off as shown in Figure 3.
3. Gently pull out plastic circuit board with battery chamber approximately 1 cm (slightly less than1/2
(See Figure 4)
4. Use pointed object such as a pen to remove old battery. Discard battery lawfully. Replace with new
1.5V DC button type LR41 or SR41,UCC392, or equivalent. Be sure battery is installed with
polarity facing up. (See Figure 5)
5. Slide battery chamber back into place and attach cover.
“
”
".)
“
+
”
LIMITED WARRANTY
Figure 2
The warranty terms here specified are valid only in Italy for Italian residents. In all other countries, the warranty will
be provided by GB the local dealer that sold you the unit, in accordance with the applicable laws.
0197
Hangzhou Sejoy Electronics &.Instruments Co., Ltd.
Building 2, No.202,Zhenzhong Road, West Lake Economy & Technology Zone 310030 Hangzhou China
Shanghai International Holding Corp.GmbH(Hamburg)
Eiffestrasse 80, 20537 Hamburg, Germany
EC
REP
M
Disposal of this product and used batteries should be carried out
in accordance with the national regulations forthe disposal of
electronic products.
The product is in compliance with the requirements
of MDD 93/42/EEC, "0197" is the identification
number of notify body.
Document No. DMT-7104-012
Version: Z
Date of Issue: 2015.05
CLEANING AND DISINFECTION
For stubborn stains, wipe the thermometer with a cloth that has been dampened with water or a
neutral detergent solution and then wring thoroughly. Finish by wiping with a soft dry cloth.
For disinfection, 70% Ethanol or Isopropyl alcohol can be used.
Observe the following to prevent damage to the thermometer.
-Do not use benzene, thinner, gasoline or other strong solvents to clean the thermometer.
-Do not attempt to disinfect the sensing section (tip) of the thermometer by immersing in
alcohol or in hot water (water over 5
0
℃
(122
℉
))
.
-Do not use ultrasonic washing to clean the thermometer.
wipe the thermometer with a soft clean cloth.
℃
/
℉
SWITCHABLE
Temperature readings are available in the Fahrenheit or Celsius scale (
℃
/
℉
; located in the upper right
corner of LCD.) With the unit off, press and hold the
On/Off
Button for approximately 4 seconds to
change the current setting.
PRECAUTION
*
The performance of the device may be degraded should one or more of the following occur:
- Operation outside the manufacturer’s s tated te mperature and humidity range.
- Storage outside the manufacturer’s s tated te mperature and humidity range.
- Mechanical shock (for example, drop test).
- Patient temperature is below ambient temperature.
*Portable and mobile RF communications can affect the device. The device needs special pre-cautions
regarding EMC according to the EMC information provided in the accompany documents.
DIGITAL THERMOMETER
This digital thermometer provides a quick and highly accurate reading of an individual's body
temperature. The digital thermometer is intended to measure the human body temperature in
regular mode orally, rectally or under the arm, and the devices are reusable for clinical or home
use on people of all ages. To better understand its functions and to provide years of dependable
results, please read all instructions first.
This appliance conforms to the following standards:
EN 12470-3 Clinical thermometers —Part 3: Performance of compact electrical thermometers
(non-predictive and predictive) with maximum device,
ISO 80601-2-56 Medical electrical equipment —Part 2-56:Particular requirements for basic safety
and essential performance of clinical thermometers for body temperature measurement,
EN 60601-1-11 Medical electrical equipment —Part 1-11: General requirements for basic safety and
essential performance – Collateral Standard: Requirements for medical electrical equipment and
medical electrical systems used in the home healthcare environment and complies with the
requirements of EN 60601-1-2(EMC), IEC/EN 60601-1(Safety) standards. And the manufacturer is
ISO 13485 certified.
PLEASE READ CAREFULLY BEFORE USING
PRODUCT ILLUSTRATION
CONTENTS
Wa r ning :
OWNER S M
ANUAL
’
1 Thermometer, 1 Owner s Manual, 1 Storage Case
’
Model:MT-31
Read instructions thoroughly before using digital thermometer.
Choking Hazard: Thermometer cap and battery may be fatal if swallowed. Do not allow children
to use this device without parental supervision.
Do not use thermometer in ear. Designed use is for oral, rectal, and armpit (axilla) readings only.
Do not place thermometer battery near extreme heat as it may explode.
Remove battery from the device when not in operation for a long time.
Note: Use of the probe cover may result in a 0.1
℃
(0.2
℉
) discrepancy from actual temperature.
It is recommended the performance should be checked every two years.
The use of temperature readings for self-diagnosis is dangerous. Consult your doctor for the
interpretation of results. Self-diagnosis may lead to the worsening of existing disease conditions.
Do not attempt measurements when the thermometer is wet as inaccurate readings may result.
Do not bite the thermometer. Doing so may lead to breakage and/or injury.
Do not attempt to disassemble or repair the thermometer. Doing so may result in inaccurate readings.
After each use, disinfect the thermometer especially in case the device is used by more than one person.
Do not force the thermometer into the rectum. Stop insertion and abort the measurement when pain is
present. Failure to do so may lead to injury.
Do not use thermometer orally after being used rectally.
For children who are two years old or younger, please do not use the devices orally.
SPECIFICATIONS
SYMBOL EXPLANATION
Measure Range:
Accuracy:
Operating mode: Direct Mode
Display:
Memory:
Battery:
Battery life:
Dimension:
Weight:
Ambient operating range:
Storage and transportation
condition:
32.0
℃-
42.9
℃
(90.0
℉
-109.9
℉
)(
℃
/
℉
chosen by manufacturer)
±
0.1
℃
(0.2
℉
) during 35.5
℃
~42.0
℃
(95.9
℉
~107.6
℉
) at
18
℃
~28
℃
(64.4
℉
~82.4
℉
) ambient operating range
±
0.2
℃
(0.4
℉
) for other measuring and ambient operating range
Procedures for determining the accuracy of the thermometer are
available upon request.
Liquid crystal display, 3 1/2 digits
For storing the last measured value
One 1.5 V DC. button battery (size LR41or SR41, UCC 392)
Approx 200 hours
13.0cm x 1.9cm x 1.0cm (L x W x H)
Approx. 12 grams including battery
Temperature:
(
)
5
℃~
40
℃
41
℉~
104
℉
Relative humidity: 15%~95%RH
Temperature:
(
)
-20
℃
~55
℃
-4
℉~
131
℉
Relative humidity: 15%~95%RH
Type BF
Classification:
Type:
Digital Thermometer (Not Predictive)
Battery Cap
O n/O ff
Button
LCD
Probe
Figure 1
Direct Current
Consult Accompanying Documents
Type BF Applied Part
Batch code
Manufacturerer
EC
REP
European Authorized Representative
Figure 4
Figure 5
Figure 3
Keep Dry
Cod. 17825 Rev.06/2017
Distribuito da / distribuited by:
FLAEM NUOVA S.p.A. - Via Colli Storici, 221
25015 S. Martino della Battaglia (BS) Italy
www.flaem.it