Calibration procedure
In order to limit parameter handling, FKG recommends reprogramming the factory settings of the memory «
M6
»
. To do
this, please refer to the programming protocol on the front page of this document.
Before using directly after purchase, whenever the motor handpiece or contra angle has been replaced, or if the motor alternates
between forward and reverse rotation outside the canal, it is recommended to calibrate the instrument in order to control and
adjust the measurements. To do this please follow the instructions below:
1. Put a commonly used file into the contra angle.
2. Turn the unit on and make sure the battery is fully charged (three bars are displayed
).
3. Select memory M6 using the Plus or Minus switch
.
4. Set the rotation mode for “Rev”:
a. Press the Select switch and select the primary function “Rotation mode”. By pressing several times, you will display the
primary functions.
b. Press the Plus or Minus switch
to set the “Rev” rotation mode.
5. Turn the unit off. To do this, simultaneously press the Select switch and the Main switch .
6. Hold down the Plus and Minus switches
and turn the unit back on.
7. When “CO-Adjst” appears in the display, press the Selct switch .
The motor will start running. Make sure there is no load on
the file.
8. When “Finished” appears in the display, the motor will stop and the calibration will be completed (press the Main switch to
exit the calibration mode).
Recommended settings for FKG instruments
Further information
Position not recommended
• Finger position
For reprocessing or complicated treatments requiring strong pressure from the operator, in order to reduce stress at the
intersection between the contra angle and the motor handle, place your fingers on the countra angle rather than the motor
handle.
Recommended position
iRace
ScoutRace
Race ISO 10
BT-Apisafe
BT-Race
Race
BioRace
Shaper
Finisher
Finisher R
DR1
DR2
Op
�
mal Speed (rpm)
Speed*
1'000
1'000
1'000
600
800
800
800
800
600
600
1'000
600
Reverse Torque (Ncm)
Rev. Torque*
1
1
1
1.5
1.5
1.5
1
1.5
1.5
1.5
1.5
1.5
Rota
�
on Mode
Rot. Mode*
Torque Slow Down
Torq. SL.D.*
Linked Func
�
on
Linked Func
�
on*
Apical Reverse or Stop
Apical ACT.*
Auto Start/Stop
Auto STT/STP*
Apical Slow Down
Apical SL.D.*
Apical Torque Reduc
�
on
Apical TRQ.D.*
M
ai
n
se
�
n
gs
Forward (Fwd*)
Su
b
-s
e
�
n
gs
These func
�
ons are not required for FKG instruments and can therefore be deac
�
vated.
Se
�
ngs
XP-endo®
D-Race
• Restore default memories
Restore the initial settings for the memories in the following way:
1. Hold down the Select switch , Plus and Minus switches
and then turn the unit on with the Main switch .
2. The “MemClear” display will appear. Press the Select switch to restore the default memories or press the Main switch
to cancel the operation.
3. Wait unit the « Finished » display appears and then press the Main switch to go to the standby display.
It is impossible to
restore the settings of a single memory.
For more information, please visit
www.fkg.ch
and/or scan the QR code above.
FKG - 2017-10 - © FKG Dentaire SA - Pictrues of our products are not contractual.
* Abbreviation of the function according to the display of the Rooter S.