1
2
4
3
2
1
3
5
4
5
Memory Number
Rotation Mode
Battery Power
Speed Setting
Torque Reverse Setting
Golden rules
Rooter S
• Before using Rooter S directly after purchase, make
sure the battery is fully charged.
• Before using directly after purchase, it is important to
calibrate the instrument. Please refer to the calibration
procedure on the back of this present document.
• Make sure the Rooter S is turned off before inserting
or removing files. Furthermore never insert or remove
files without holding down the file release or insertion
switch, this will damage the chuck.
• FKG instruments are used in continuous rotation and
do not require the OTR function.
Quick start instructions
Plus or Minus
Switch
Select
Switch
Programming protocol
The factory settings can be changed according to your
wishes and/or according to the FKG instrumentation
table (see overleaf). To do this, please use the following
programming protocol:
1. Press the Main switch to turn Rooter S on. The
standby display will appear.
2. Select memory number (for example M6): press Plus
or Minus switch
.
3. Changing a parameter of a primary function (speed,
torque, rotation mode):
a. Press the Select switch to choose one of the
primary functions. By pressing several times, you
will display the primary functions.
b. Press the Plus or Minus switch
to change the
desired setting.
4. Changing a sub-setting of the same memory number:
a. Hold down the Select switch for at least 1
second.
b. Press the Select switch to display the various
sub-settings.
c. Change the setting; press the Plus or Minus switch
.
5. Changing a parameter of an optional setting (volume,
right or left handed display, return to standby time,...):
a. Hold down Select switch and then press the
switch .
b. Press the Select switch to select one of the
settings.
c. Press the Plus or Minus switches
to change
the setting.
d. Press the Main switch to return to the standby
display.
File release
or insertion
switch
Main switch
Turn power on:
Turn power off: +
LCD display -
Standby display