FiT333
:
333-15Y07526 2015-11-12
9
2. Quick Start: - Under this mode, the device allow to test without initialize
sensor condition. Eg: temperature & residual alcohol.
It can ready to test within 1 second.
Repeat the 1.1 – 1.8 to do the quick breath alcohol test.
3. System Setting
(Available for Service Center only):
Select the “ ” menu, and input the password (The default password is nothing, only press the OK key to
enter into the system setting). Then, shows the sub-menu as following: Set unit, “Set Date & Time”, “Password”,
“System Information”, “Advanced Setting”.
Caution:
The calibration is already been done before shipping, user alters the setting parameters will override
all calibration data. Only FiT authorized technician is allowed to modify the calibration data.
①
Set Unit
Set the unit to Permillage (
‰
B
AC),
Percentage (
%
B
AC) and mg/L, and mg/100ml. as picture:
②
Set Date & time: under this model user can
input password
to set
the new date & time ;
“calibration interval” (default 365days) for
Calibration Reminder / Automatically Stops Testing When Unit is due for
Calibration
;
The calibration interval base day is
“Update 0.02%
Calibration date”
Input the real year/month/date/ hour/minute/second
by turn and press
"
OK
"
key to save and return back