8.11
Holding the dispense gun over test sheet
press the foot pedal once to activate the
dispense cycle and observe fluid flow.
Wait until to the fluid comes to a
complete stop before starting
another dispense.
8.12
If fluid is flush at the end of the dispense tip
proceed to 8.20.
8.13
If fluid is oozing proceed to 8.14. If fluid is
recessed proceed to 8.17.
8.14
Tap/click the AUTO(Manual)
button to MANUAL(Auto)
and then tap/click the
EXTEND(Retract) button to
RETRACT(Extend).
8.15
Press the foot pedal as many
times as needed to stop the
fluid flow. If you go too far
and suck fluid into the tip
tap/click
RETRACT(Extend)
to
EXTEND(Retract)
and repeat
until fluid is settled at the tip.
8.16
Return to procedural step 8.9, increasing and
decreasing
Back-off steps
until your dispense
tip is flush at the end of dispense. If flush
proceed to 8.20.
Interna00 1 508-435-2115
www.fishmancorp.com
EXTEND
(Retract)
RETRACT
(Extend)
AUTO
(Manual)
MANUAL
(Auto)
RETRACT
(Extend)
EXTEND
(Retract)