Measuring
7.3
Making a Measurement
To make a measurement, place the probe at a right angle on the specimen
surface (
Page 34). The probe can be lifted off after the measurement ac-
quisition, i.e., after the reading appears on the display. The instrument is
ready to make measurements.
Observe the following during the measurement:
Measurements should be made within the reference area.
To avoid erroneous readings, do not allow the probe to hover above
the specimen.
To obtain a correct air value, the distance to the specimen should be at
least 25 mm.
To allow sufficient time for a measurement accept, the time between
individual measurements must be greater than 0.5 seconds.