Fisher Scientific Mini300V Plus Instruction Manual Download Page 3

 

 

 

 

 

 

 

Warning

   

The MINI300V PLUS Power Supply has been thoroughly tested and found to 

comply within the limits of CE regulation.    It has been manufactured using the 

latest technology and does not require maintenance.    When used correctly 

this unit poses no particular health risk, although it can deliver dangerous 

voltage levels if used incorrectly.    Accordingly, this power supply must only be 

operated by fully qualified personnel adhering to the guidelines laid out within 

this instruction manual.    Although this power supply is equipped with all 

necessary safety features against abuse and accidental failure, caution should 

always be exercised when working with high voltage equipment.    Any 

individual intending to use this instrument should read the entire manual 

thoroughly before operation. 

 

1.  Read the instruction manual thoroughly before use. 

2.  Never touch the power outlets with any conductive object (e.g. naked 

metal wire) other than properly insulated power supply cables. 

3.  Do not spill liquid or insert metal objects inside the power supply. 

4.  Never block the ventilation holes or place the unit in any enclosure unless 

there is adequate ventilation; never expose the power supply to a direct 

heat source. 

5.  Never touch any part of the power supply assembly (i.e. power supply, 

cables or electrophoresis tank) before switching OFF the power supply. 

6.  Never manipulate with wet hands. 

7.  Do not connect to ground any of the power outputs or the buffer within the 

electrophoresis tank; the power outputs should be only connected to an 

insulated electrophoresis tank equipped with a safety cover. 

8.  Do not connect any power supplies in series or in parallel. 

9.  Never open the back plate or remove the cover, otherwise an electric 

shock may result.    Repairs should only be made by the manufacturer or a 

service technician authorised by the manufacturer. 

10.  Never use this power supply if the safety cover is not in position correctly. 

11.  Do not use the unit if there is any sign of damage to the external tank or 

cover.    Contact the manufacturer or supplier immediately to replace or 

Summary of Contents for Mini300V Plus

Page 1: ...1 Mini300V Plus Power Supply Instruction manual Mini300V Plus Version 01 May 2012...

Page 2: ...2 Packing list Mini300V Plus 1x Mini300V Plus Power Supply 1x Power Cord 1x Instruction Manual Signed by Date...

Page 3: ...operly insulated power supply cables 3 Do not spill liquid or insert metal objects inside the power supply 4 Never block the ventilation holes or place the unit in any enclosure unless there is adequa...

Page 4: ...supply may only be installed and operated only under the following environmental conditions 1 For indoor use only 2 Relative humidity 95 3 Atmospheric pressure 75 kPa 106 kPa 4 Altitude 2000 metres 5...

Page 5: ...ation 8 Section 3 Installation Instructions 9 Section 4 Operation Instructions 9 4 1 Control interface 9 4 2 Start the operation 10 4 3 To use the constant current or voltage mode 11 4 4 To adjust the...

Page 6: ...6 Section 1 Introduction 1 1 Overview Front view Rear view Main power socket Fuse holder ON OFF Switch Fan...

Page 7: ...n or timed setting up to 999 minutes The MINI300V PLUS s user friendly interface is easily adjustable in 1V and 1mA increments making it perfect for separations where precise settings are required whi...

Page 8: ...3 digit LED Programmable Methods Automatic recovery after power failure Safety features No load detection over voltage current temperature protection maximum power output detection shrouded plugs and...

Page 9: ...ventilated and away from sources of direct heat Section 4 Operation Instructions 4 1 Control interface There are six buttons and four LED indicators on the faceplate The LED indicates the operation st...

Page 10: ...alue For example to operate under constant voltage adjust the current to the maximum output of 400mA before running the power supply with the voltage set at the desired output setting 1 Place the powe...

Page 11: ...voltage or current in 1V or 1mA increments or decrements to the required output value 6 After adjusting your constant mode output value e g voltage set the other parameter to the maximum output e g cu...

Page 12: ...e instructions in this manual accordingly Some suggestions for troubleshooting are given below If these suggestions fail to resolve the problem contact the Fisher Scientific distributor in your region...

Page 13: ...esis tank Error message Current exceeds the maximum output for the power supply 400mA Check if the buffer concentration or molarity is appropriate Excessive buffer concentration or molarity may increa...

Page 14: ...ch at the rear of power supply and detach the power cord 2 Open the fuse compartment located inside the Power Entry Module by inserting a small flat bladed screwdriver into the slot below the ON OFF s...

Page 15: ...300V PLUS Power Supply uses all solid state components and should require no maintenance or recalibration under normal use If the unit must be returned for repair contact your local authorised Fisher...

Page 16: ...16 Section 6 Ordering Information Cat No Description FB79520 Fisher Scientific MINI300V PLUS 300V 400mA 60W Power Supply 110 240 VAC...

Page 17: ...by improper operation Units where repair or modification has been performed by anyone other than Fisher Scientific or an appointed distributor or representative are no longer under warranty from the...