Isotemp Water Bath Quick Start Guide

This quick start guide is intended for initial start up only. For all other procedures you must refer to the manual. Also, if any 
of these steps are not clear refer to the manual before proceeding.

Filling 

(Only approved fluids are Filtered/Single Distilled Water and De-ionized Water)

Ensure the reservoir drain port is closed and that all plumbing connections are securely plumbed or capped. Also ensure any residue is 
thoroughly removed from the reservoir before filling.

To avoid spilling, place your samples/trays into the bath before filling. 

Slowly

 fill the reservoir. When adding, point the opening of a container away from yourself. 

For Shaking Water Baths, fill the reservoir with a minimum of one inch of water and the maximum level lower than the tray shaft fitting. 
Make allowance for water splashing from tray oscillations.

For GPD Baths, fill the reservoir with a minimum of one inch of water and a maximum level one inch lower than the tank upper surface.

Starting

Do not run the bath until fluid is added to the reservoir. Have extra fluid on hand. If the bath does not start refer to the manual.

 Place the circuit protector located on the rear of the bath to the 

I

 position.

 The screen will momentarily display 

 and then the home screen appears displaying the set point and the audible alarm status. 

The set point is the desired reservoir fluid temperature.

30.0°C

        appears  on  Shaking                  

 

          Water Baths only.

Page 1 of 2

20.1°C

30.0°C

Press 

 to start the bath and display the fluid temperature. If the reservoir fluid 

temperature is below the set point the heater 

 will start.

Summary of Contents for Isotemp

Page 1: ...User s Manual Manual P N U01318 Rev 11 05 2015 Isotemp Shaking and General Purpose Water Baths...

Page 2: ...irements 3 2 Approved Fluids 3 3 Filling Requirements 3 6 Filling with Thermal Lab Beads 3 6 Draining 3 7 Accessories 3 7 Drain Kit Accessory 3 7 Chapter 4 Operation 4 1 Controller 4 1 Start Up 4 2 Ch...

Page 3: ...Isotemp Fisher Scientific...

Page 4: ...e reservoir with a minimum of one inch of water and the maximum level lower than the tray shaft fitting Make allowance for water splashing from tray oscillations For GPD Baths fill the reservoir with...

Page 5: ...n view change the speed value with the bath running or not From the home screen press the soft key below to display Press and hold the arrow keys to bring up the desired value The values are 0 and 30...

Page 6: ...hstens bis unterhalb der Halterung f r den Tablettschaft Lassen Sie etwas Platz f r Spritzwasser durch die Schwingungen des Tabletts F r GPD B der f llen Sie den Beh lter mindestens bis zu einer F llh...

Page 7: ...eschwindigkeitseinstellung l sst sich sowohl bei eingeschaltetem als auch bei ausgeschaltetem Bad anzeigen ndern Vom Einstiegsbildschirm aus dr cken Sie die Schaltfl che unten zur Anzeige von Dr cken...

Page 8: ...te les claboussures dues aux oscillations du plateau Pour les bains GPD remplissez le r servoir d au moins 2 5 centim tres 1 pouce d eau et assurez vous que le niveau maximum se trouve 2 5 centim tres...

Page 9: ...icher Appuyez sur les touches fl che et maintenez les enfonc es jusqu atteindre la valeur d sir e Les valeurs sont 0 et de 30 200 Une fois que la valeur d sir e est affich e appuyez sur ou pour enregi...

Page 10: ...inferior al ajuste del eje de la bandeja Tenga en cuenta las salpicaduras de agua procedentes de las oscilaciones de la bandeja En el caso de ba eras GPD llene el dep sito con un m nimo de dos cent m...

Page 11: ...echa para seleccionar el valor del punto de ajuste deseado Una vez que aparezca resaltado el punto de ajuste pulse la siguiente tecla multifunci n para guardar los cambios y volver a la pantalla de in...

Page 12: ...legada de gua e o n vel m ximo mais baixo que a conex o do eixo da bandeja Deixe uma folga para respingos de gua durante as oscila es da bandeja Para Banhos GPD encha o reservat rio com no m nimo uma...

Page 13: ...trar o valor de ajuste desejado A varia o de 5 C a 100 C Quando o valor desejado for exibido neste caso 50 C pressione a tecla program vel abaixo para salvar a altera o ou pressione a tecla program ve...

Page 14: ...does not have handles Take into account its weight when unpacking and transporting We recommend two people lift heavier baths from the bottom After sale Support Fisher Scientific is committed to cust...

Page 15: ...istributor s sole option any Product which is defective in material or workmanship Distributor expressly disclaims all other warranties whether expressed implied or statutory including the warranties...

Page 16: ...l result in death or serious injury WARNING indicates a potentially hazardous situation which if not avoided could result in death or serious injury CAUTION indicates a potentially hazardous situation...

Page 17: ...removed Operating the baths at high temperature will cause condensation on the underside of the cover Remove the cover with care so the water falls back into the reservoir and not on the bath surface...

Page 18: ...is designed for operation with water The body is made from galvanized steel and is painted for added protection A drain is located at the far left hand end of the bath A stainless steel gable cover r...

Page 19: ...1200 1500 Bath Volume liters 15 27 Overall Bath Dimensions L x W x H cm inches 39 4 x 63 2 x 24 9 15 5 x 24 9 x 9 8 39 4 x 93 8 x 24 9 15 5 x 36 9 x 9 8 Bath Work Area Dimensions L x W x H cm inches 2...

Page 20: ...bient Temperature C Range F 15 to 45 59 to 113 Stability 37 C 0 1 Uniformity 37 C 0 2 Heating Output watts 200 300 300 800 Bath Volume liters 2 2 5 10 Overall Bath Dimensions L x W x H cm inches 23 0...

Page 21: ...to 100 9 to 212 Ambient Temperature C Range F 15 to 45 59 to 113 Stability 37 C 0 1 Uniformity 37 C 0 2 Heating Output watts 1200 1200 300 and 800 Bath Volume liters 20 28 5 and 10 Overall Bath Dimens...

Page 22: ...tion The bath is designed for continuous operation and for indoor use Never place the bath in a location where excessive heat moisture or corrosive materials are present Ventilation No special ventila...

Page 23: ...ion is provided The circuit protector on the rear of the bath is designed to protect the bath s internal components Note If the circuit protector activates allow the temperature to cool before resetti...

Page 24: ...mmended to regularly drain and replace fluid especially when higher temperatures are utilized above 40 C Draining and drying the bath after each use is recommended as leaving standing water for any pe...

Page 25: ...vent electrical leaks to ground through the recirculating fluid De ionized water is in an unbalanced state and will leach the missing ions from the materials it comes in contact with The aggressive na...

Page 26: ...e 25 1 0 Sodium 20 0 3 Sulfate 25 1 Hardness 17 0 05 Total Dissolved Solids 50 10 Other Parameters pH 6 5 8 5 7 8 Resistivity 0 01 0 05 0 1 M cm compensated to 25 C Unfavorably high total ionized soli...

Page 27: ...mum of one inch of water and the maximum level lower than the tray shaft fitting For General Purpose Baths fill the reservoir with a minimum of one inch of water and a maximum level one inch lower tha...

Page 28: ...Hose 3 8ID x 5 8OD Figure 3 4 Drain Kit Accessory CAUTION Accessories Only use the accessories supplied by Fisher Scientific Figure 3 3 GPD Bath Drain Hose Remove the drain hose from its housing and...

Page 29: ...3 8 Isotemp Fisher Scientific Chapter 3 Installation...

Page 30: ...ss the two navigation arrows to move through the controller displays and to adjust values Press a soft key to select additional displays Indicates the heater is on The icon flashes when the reservoir...

Page 31: ...glass vessels are secure Unrestrained vessels may shift and break when shaking is engaged Flask clips and trays are available from Fisher Scientific Do not run the bath until fluid is added to the res...

Page 32: ...the soft key below to save the change or press the soft key below x to not save the change Note After a 30 second delay if neither key is pressed the display returns to the home screen any change is n...

Page 33: ...the audible alarm status The top indicates it is enabled the bottom indicates disabled Press to toggle between enabled disabled Use the arrow keys to highlight another setting or press to return to th...

Page 34: ...The top indicates it is enabled the bottom is disabled Press to toggle between enabled disabled Press to return to the home screen When enabled and the reservoir level nears empty the bath will shut...

Page 35: ...meter and press again Then press and hold the arrow keys to change the value Once the desired value is displayed press If desired press the arrow keys to highlight another preset parameter or press to...

Page 36: ...th shut down place the circuit protector located on the rear of the bath to the O position To disconnect the bath remove the electrical plug from its power source With the bath running and after 30 se...

Page 37: ...4 8 Isotemp Fisher Scientific Chapter 4 Operation...

Page 38: ...cloth and warm water only After time the circulating bath s stainless steel surfaces may show spots and become tarnished Normal stainless steel cleaners can be used Clean the reservoir and built in c...

Page 39: ...ing the top right panel 2 Swing panel assembly open Do not operate the bath with the panel open 3 Rotate disc by grasping aluminum components of the eccentric only one turn clockwise and verify that d...

Page 40: ...high temp protection limit exceeded allow bath to cool down restart the bath if the HTC fault cannot be cleared the bath must be serviced by an authorized Fisher Scientific Temperature Control Service...

Page 41: ...fy the set point Low end temperature set points require supplemental cooling GPD 02 baths require insulation especially in the lid for operation above 90 C Ensure bath installation complies with the s...

Page 42: ......

Reviews: