Fisher Scientific Fisherbrand accuSpin Micro Series Instruction Manual Download Page 31

31 / 89

Technical Specifications    

17R Series Centrifuges – PCR 4 x 8 Rotor

Voltage

230 V

120 V

Maximum Speed

13 300

13 300

Maximum RCF-Value

13 100

13 100

Acceleration / Braking Time

9 s / 12 s

9 s / 12 s

Min. Temperature at max. Speed, Ambient 
Temperature of 23 °C

≤ 0 °C

≤ 0 °C

21R Series Centrifuges – PCR 4 x 8 Rotor

Voltage

230 V

120 V

Maximum Speed

14 800

14 800

Maximum RCF-Value

16 200

16 200

Acceleration / Braking Time

11 s / 13 s

11 s / 13 s

Min. Temperature at max. Speed, Ambient 
Temperature of 23 °C

≤ 0 °C

≤ 0 °C

Table 32: 

PCR 4 x 8 Rotor Performance Data

Accessories

Description

Article 
No� 

Rotor 
Capacity

Max� Tube 
Dimensions

ClickSeal biocontainment lid

75003410

-

-

Replacement O-ring set for ClickSeal lid – 
75003410

75003405

-

-

Table 33: 

Accessories PCR 4 x 8 Rotor

1. 6. 7.   Hematocrit Rotor

For more information on the hematocrit rotor refer to the separate rotor manual.

Summary of Contents for Fisherbrand accuSpin Micro Series

Page 1: ...Fisherbrand accuSpin Micro Series Instruction Manual 50165241 b 2022 03...

Page 2: ...Micro 17 12 Fisherbrand accuSpin Micro 17R 13 Fisherbrand accuSpin Micro 21 14 Fisherbrand accuSpin Micro 21R 15 1 3 1 Directives and Standards 16 1 4 Mains Supply 18 1 5 Refrigerants 19 1 6 Rotor Spe...

Page 3: ...CF Value Explained 46 3 5 Entering Centrifugation Parameters 47 3 5 1 Selecting Speed or RCF Value 47 3 5 2 Pre selecting Speed 47 3 5 3 Pre selecting the RCF Value 48 3 5 4 Pre selecting Run Time 48...

Page 4: ...ection 58 4 5 Decontamination 58 4 6 Autoclaving 59 4 7 Service 59 4 8 Lifetime 60 4 9 Shipping 60 4 10 Storage 60 4 11 Disposal 61 5 Troubleshooting 62 5 1 Mechanical Emergency Lid Release 62 5 2 Ice...

Page 5: ...aboratories by trained personal Signal Words and Symbols Signal Word Degree of Hazard WARNING Indicates a hazardous situation that if not avoided could result in death or serious injury CAUTION Indica...

Page 6: ...y instructions The centrifuge is to be used for its intended use only Improper use can cause damages contamination and injuries with fatal consequences The centrifuge must be operated by trained perso...

Page 7: ...and any local regulations When centrifuging microbiological samples from the Risk Group II according to the Laboratory Biosafety Manual of the World Health Organization WHO aerosol tight biological s...

Page 8: ...e occurs Do not open the centrifuge before the rotor has stopped spinning Do not touch a spinning rotor Open the centrifuge only if the rotor has stopped spinning Never use your hands or tools to stop...

Page 9: ...it is running The centrifuge housing is not to be opened by the operator CAUTION Due to air friction sample integrity may be affected The temperature of the rotor may rise significantly while the cen...

Page 10: ...main switch Pull out the power supply plug In an emergency disconnect the power supply Make sure that the main switch and power supply plug can be accessed easily when setting up the centrifuge The g...

Page 11: ...rbrand accuSpin Micro 21R 120 V 10 60 Hz 75003248 Fisherbrand accuSpin Micro 21 230 V 10 50 60 Hz 75003249 Fisherbrand accuSpin Micro 21R 230 V 10 50 60 Hz Table 1 List of Centrifuges 1 2 List of Roto...

Page 12: ...at Dissipation 0 15 kW h Environmental Conditions For Storage and Shipping Temperature 2 C to 50 C For Operation Use in interior spaces Altitudes of up to 3000 m above sea level Temperature 2 C to 40...

Page 13: ...ion 0 25 kW h Environmental Conditions For Storage and Shipping Temperature 2 C to 50 C For Operation Use in interior spaces Altitudes of up to 3000 m above sea level Temperature 2 C to 40 C Max relat...

Page 14: ...ation 0 2 kW h Environmental Conditions For Storage and Shipping Temperature 2 C to 50 C For Operation Use in interior spaces Altitudes of up to 3000 m above sea level Temperature 2 C to 40 C Max rela...

Page 15: ...tion 0 3 kW h Environmental Conditions For Storage and Shipping Temperature 2 C to 50 C For Operation Use in interior spaces Altitudes of up to 3000 m above sea level Temperature 2 C to 40 C Max relat...

Page 16: ...lectronic equipment EN 61010 1 3 1 Edition EN 61010 2 020 3rd Edition EN 61010 2 011 2nd Edition EN 61010 2 101 3rd Edition EN 61326 1 Class B EN ISO 14971 ISO 13485 North America FDA listed Product c...

Page 17: ...nterference to radio communications However there is no guarantee that interference will not occur in a particular installation If this equipment does cause harmful interference to radio or television...

Page 18: ...Spin Micro 17R 230 V 10 50 60 Hz 1 9 A 320 W 4 A 2 pole circuit breaker 16 A 75002463 Fisherbrand accuSpin Micro 17R 120 V 10 60 Hz 3 9 A 330 W 8 A 2 pole circuit breaker 15 A 75003246 Fisherbrand acc...

Page 19: ...in Micro 17R R 134a 0 26 kg 21 bar 1430 0 37 t 75002463 Fisherbrand accuSpin Micro 17R R 134a 0 26 kg 21 bar 1430 0 37 t 75003247 Fisherbrand accuSpin Micro 21R R 134a 0 26 kg 21 bar 1430 0 37 t 75003...

Page 20: ...rature 121 C Table 11 General Technical Data 24 x 1 5 2 0 mL Rotor Rotor Performance Data 17 Series Centrifuges 24 x 1 5 2 0 mL Rotor Voltage 230 V 120 V Maximum Speed 13300 13300 Maximum RCF Value 17...

Page 21: ...Acceleration Braking Time 12 s 13 s 12 s 13 s Min Temperature at max Speed Ambient Temperature of 23 C 0 C 0 C Table 12 24 x 1 5 2 0 mL Rotor Performance Data Accessories Description Article No Rotor...

Page 22: ...15 General Technical Data Dual Row 18 x 2 0 0 5 mL Rotor Rotor Performance Data 17 Series Centrifuges Dual Row 18 x 2 0 0 5 mL Rotor Voltage 230 V 120 V Maximum Speed 13300 13300 Maximum RCF Value 16...

Page 23: ...f 23 C 0 C 0 C 21R Series Centrifuges Dual Row 18 x 2 0 0 5 mL Rotor Voltage 230 V 120 V Maximum Speed 14800 14800 Maximum RCF Value 20800 20800 Acceleration Braking Time 11 s 13 s 11 s 13 s Min Tempe...

Page 24: ...General Technical Data 36 x 0 5 mL Rotor Rotor Performance Data 17 Series Centrifuges 36 x 0 5 mL Rotor Voltage 230 V 120 V Maximum Speed 13300 13300 Maximum RCF Value 15600 15600 Acceleration Brakin...

Page 25: ...Temperature of 23 C 0 C 0 C 21R Series Centrifuges 36 x 0 5 mL Rotor Voltage 230 V 120 V Maximum Speed 14800 14800 Maximum RCF Value 19300 19300 Acceleration Braking Time 9 s 11 s 9 s 11 s Min Tempera...

Page 26: ...erature 121 C Table 23 General Technical Data 10 x 5 mL Rotor Rotor Performance Data 17 Series Centrifuges 10 x 5 mL Rotor Voltage 230 V 120 V Maximum Speed 13300 13300 Maximum RCF Value 16414 16414 A...

Page 27: ...ies Centrifuges 10 x 5 mL Rotor Voltage 230 V 120 V Maximum Speed 14800 14800 Maximum RCF Value 20326 20326 Acceleration Braking Time 12 s 13 s 12 s 13 s Min Temperature at max Speed Ambient Temperatu...

Page 28: ...able 27 General Technical Data PCR 8 x 8 Rotor Rotor Performance Data 17 Series Centrifuges PCR 8 x 8 Rotor Voltage 230 V 120 V Maximum Speed 13300 13300 Maximum RCF Value 13800 13800 Acceleration Bra...

Page 29: ...t Temperature of 23 C 0 C 0 C 21R Series Centrifuges PCR 8 x 8 Rotor Voltage 230 V 120 V Maximum Speed 14800 14800 Maximum RCF Value 17100 17100 Acceleration Braking Time 7 s 9 s 7 s 9 s Min Temperatu...

Page 30: ...e 31 General Technical Data PCR 4 x 8 Rotor Rotor Performance Data 17 Series Centrifuges PCR 4 x 8 Rotor Voltage 230 V 120 V Maximum Speed 13300 13300 Maximum RCF Value 13100 13100 Acceleration Brakin...

Page 31: ...aximum Speed 14800 14800 Maximum RCF Value 16200 16200 Acceleration Braking Time 11 s 13 s 11 s 13 s Min Temperature at max Speed Ambient Temperature of 23 C 0 C 0 C Table 32 PCR 4 x 8 Rotor Performan...

Page 32: ...report it to the carrier and request a damage inspection Important Failure to request an inspection of damage within a few days after receipt of shipment absolves the carrier from any liability for d...

Page 33: ...o compensate for an uneven surface Do not operate the centrifuge on carts or stand alone shelving that could move during operation or is inadequately sized for the centrifuge Be able to hold the weigh...

Page 34: ...34 89 Transport and Set Up Figure 1 Safety Zone...

Page 35: ...t the centrifuge the centrifuge door is closed CAUTION Hands can be squeezed by an open centrifuge door Always close the centrifuge door before you transport the centrifuge Before transporting a rotor...

Page 36: ...h 2 Make sure that the power supply cable specification agrees with the safety standards of your country 3 Make sure that the voltage and frequency are the same as the figures on the nameplate 4 Make...

Page 37: ...s You can pre temper the centrifugation chamber and the empty rotor before the centrifugation run starts using SNOWFLAKE button NOTICE This function is only available on refrigerated centrifuges PULSE...

Page 38: ...h for balancing potential voltage discrepancies in the power grid After pressing the mains switch the display therefore may still flash up to 10 seconds 3 2 3 Audible Alarm Error Accompanying all erro...

Page 39: ...the centrifuge lid Close the centrifuge lid by pressing down on it lightly in the middle or on both sides The locking mechanism engages to close the lid safely The lid should audibly click into place...

Page 40: ...fuge spindle The two bars in the labeling on the upper side of the rotor a must be aligned with the retaining pin b of the centrifuge spindle a b 3 Let the rotor slide down slowly 4 Insert the Allen w...

Page 41: ...be hot 55 C Be aware of this risk and proceed carefully when you change a rotor after a run or wait until the motor has cooled down Before installing a rotor Remove any dust foreign objects or residue...

Page 42: ...1 Put the rotor lid on the rotor nut 2 Push the rotor lid down until you see and hear the lock click If the lid does not close at all or only by force make sure that the sealing rings are correctly i...

Page 43: ...otor Lid Seals CAUTION Loose parts may damage the centrifuge When operating the rotor without the lid the seals are not fixed in their position and can damage the centrifuge Tube Caps Always close the...

Page 44: ...rrosion 2 Inspect the centrifugation chamber and drive shaft for damage such as cracks scratches or traces of corrosion 3 Check the suitability of the rotor and other used accessories against the Chem...

Page 45: ...um Loading Each rotor is designed to run with its maximum load at maximum speed The safety system of the centrifuge requires that the rotor is not overloaded The rotors are designed to work with subst...

Page 46: ...maximum RCF value is related to the maximum radius of the tube opening Remember that this value is reduced depending on the tubes buckets and adapters used This can be accounted for in the calculatio...

Page 47: ...3 5 2 Pre selecting Speed 1 Enter the desired value by pressing the MINUS PLUS buttons repeatedly until the desired value shows You can adjust the speed in 100 rpm increments 2 Press START to confirm...

Page 48: ...entrifuge radius of the 24 x 1 5 2 0 mL Rotor 75003424 Refer to RCF Value Explained on page 46 for further information NOTICE Avoid if possible speed ranges close to natural resonances of the system R...

Page 49: ...etermine the sample temperature in C Proceed as follows 1 Enter the desired value by pressing the MINUS PLUS buttons under the right display repeatedly until the desired value shows You increase or de...

Page 50: ...the activated pre temp function 2 Enter the desired value by pressing the MINUS PLUS buttons under the right display repeatedly until the desired value shows You increase or decrease the temperature...

Page 51: ...fuge Refer to Safety Zone on page 34 Persons and hazardous substances must be kept out of this safety zone while centrifuging Once the main switch has been turned on the rotor has been properly instal...

Page 52: ...LSE spinning will start and continue until you release the key The centrifuge accelerates to and brakes at maximum power The pre selected value is ignored NOTICE The centrifuge accelerates to maximum...

Page 53: ...p detail Large seal in outer groove of rotor body bottom detail small seal in groove of rotor collar 1 Lubricate all seals 2 Press the large seal in the outer groove of the rotor body top detail 3 Pre...

Page 54: ...id Quick Test As a quick test it is possible to test the aerosol tightness of fixed angle rotors using the following process 1 Lubricate all seals lightly Always use the supplied grease when lubricati...

Page 55: ...oroughly cleaning the rotors they must be inspected for damage wear and corrosion Make sure that sealing rings are still smooth not brittle nor otherwise damaged Some sealing rings are not autoclavabl...

Page 56: ...re of 50 C If drying boxes are used the temperature must never exceed 50 C Higher temperatures could damage the material and shorten the lifetime of the parts 8 Inspect the rotor and accessories for s...

Page 57: ...ifuge to the edge of the table 2 Pull the clip below the sucking grid and remove the filter unit completely by pulling it down 3 Remove the cumulated dust with a soft cloth 4 Inserting the filter agai...

Page 58: ...agent Observe the safety precautions and handling instructions for the disinfection agents used 4 5 Decontamination You are responsible that the level of decontamination is achieved according to your...

Page 59: ...ICE No chemical additives are permitted in the steam 4 7 Service Fisher Scientific recommends having the centrifuge and accessories serviced once a year by an authorized service technician The service...

Page 60: ...on with a decontamination certificate WARNING Before shipping the centrifuge and accessories you must clean and if necessary disinfect or decontaminate the full system If you are not sure consult with...

Page 61: ...com centrifuge For the countries of the European Union the disposal is regulated by the European Union s Waste Electrical Electronic Equipment WEEE Directive 2012 19 EC Mind the information on transpo...

Page 62: ...The braking process lasts much longer than usual Proceed as follows 1 Wait until the rotor has stopped This can take several minutes Use the view port for visual confirmation 2 Pull out the power supp...

Page 63: ...o not use any sharp tools aggressive liquids or fire to fasten the melting process If necessary use warm water to speed up the melting process 4 Remove the water from the centrifugation chamber 5 3 Tr...

Page 64: ...ation to standstill Centrifuge lid was opened manually during the run Close centrifuge lid immediately Rotor stops with deceleration to standstill For further centrifugation you have to switch the ins...

Page 65: ...respectively the filter unit of the cooled centrifuge After approx 60 minutes you can restart the instrument Observe the maximum permissible environmental temperature If the error message still shows...

Page 66: ...technician Table 34 Troubleshooting 5 3 8 Information for the Customer Service If you need to contact customer service please provide the order no and the serial no of your device This information can...

Page 67: ...ptoethanol S S U S M S S U S S U S S S S S S U S S S S S S A cetaldehyde S U U M U M U U U M M M S U S U A cetone M S U U S U M S S U U S U S U U U S S U U S M M S U U A cetonitrile S S U S M S S S U...

Page 68: ...U U S M U U S S U S U S U A mmonium A cetate S S U S S S S S S S S S S U S S S S S S S S S S A mmonium C arbonate M S U S S S S S S S S S S S U U S S S S S S M S S S A mmonium H ydroxide 10 U U S U S...

Page 69: ...S S S S S M S S S S S S M S S S A mmonium S ulfate U M S S S U S S S S S S S S S S S S S S S U S S U A myl A lcohol S M U S S M S M S S S S M U S M A niline S S U U S U S M S U U U U U U U S M U U S S...

Page 70: ...d U S S M S S U S S S S S S S S S U S S S S S S S S S S C esium A cetate M S S S S S S S S S S S S S S S S M S S S C esium B romide M S S S S S S S S S S S S S S S S S S M S S S C esium C hloride M S...

Page 71: ...romic A cid 50 U U U U U S U U S M U M S S U M S U M S C resol M ixture S S U S S U U U U U U U U U S S S S U S C yclohexane S S S S S S U S U S S U U U M S M U M M S U M M U S D eoxycholate S S S S S...

Page 72: ...nate S S U S S S S S U S U S U S S S M S S S S S S D imethylsulfoxide S S U U S S S S U S S U S U U S S U U S S S S U U D ioxane M S U U S S M M S U U S U M U U M M M U S S S S U U F erric C hloride U...

Page 73: ...S U U S M M S U U E thyl A lcohol 50 S S S S S S M S S S S S U S U S S S S S S S S M S M U E thyl A lcohol 95 S S S U S S M S S S S S U S U S S S M S S S U S M U E thylene D ichloride S U U S M U U S...

Page 74: ...ofluoric A cid 10 U U U M U U U S S M U S S S S M S U U U H ydrofluoric A cid 50 U U U U U U U U U S U U U S S M M S U U U M H ydrochloric A cid conc U U U U U U M U M U U M U U U S U S U U U F ormald...

Page 75: ...mo S ol S S S S S S S S S S S S S S S S S S S S S H exane S S S S S S S S U S U M U S S U S S M S U S S U S I sobutyl A lcohol M U S S U S U S S M S S S S S S S S I sopropyl A lcohol M M M U S S S S S...

Page 76: ...S S S S S S S S S S S S S S S S S S S U S S S P otassium H ydroxide 5 U U S S S S M S S S S S U S S S S S S S M U M S U P otassium H ydroxide conc U U M U M M S S U M U U U S M M U U U U P otassium P...

Page 77: ...n Viton K erosene S S S S S S U S M U S U M M S M M M S S U S S U S S odium C hloride 10 S S S S S S S S S S S S S S S S S S M S S odium C hloride sat d U S U S S S S S S S S S S S S S M S C arbon T e...

Page 78: ...S S S S S S S S S S S S S S S S S S S M S S S M ercaptoacetic A cid U S U S M S S M S U U U U S U U S M S U S S S S M ethyl A lcohol S S S U S S M S S S S S U S U M S S S S S S S M S M U M ethylene C...

Page 79: ...M S S M S M S S S L actic A cid 20 S S M S M S S S S S S S M S M S S S N B utyl A lcohol S S U S S M U S M S S S S M M S M S S N B utyl P hthalate S S U S S S S U U S U U U M U U S U S M M S U S N N D...

Page 80: ...S S S S S S S S S S S S S odium D odecyl S ulfate S S S S S S S S S S S S S S S S S S S S S S S S S odium H ypochlorite 5 U U M S S M U S S M S S S M S S S S M S S S M U S M S S odium I odide M S S S...

Page 81: ...kel S alts U S S S S S S S S S S S S S S S S S S M S S S O ils P etroleum S S S S U S S S S U U M S M U U S S S U S S S S O ils O ther S S S M S S S S U S S S S U S S S S S S M S O leic A cid S U S S...

Page 82: ...U M U U U U U S U M U U S U U U M S P hosphoric A cid 10 U U M S S S U S S S S U S S S S S S S S S U M U S S P hosphoric A cid conc U U M M U S M S U U M M S S S M S M S U M U S P hysiologic M edia S...

Page 83: ...S S S S S S S S S S S S S S S S S S S S S S ucrose A lkaline M S S S S S S S S S S S U S S S S S S S S M S S S S ulfosalicylic A cid U U S S S S S S S S U S S S S S S S S S U S S S N itric A cid 10 U...

Page 84: ...S S S S M S M U U M M S ulfuric A cid 10 M U U S S U U S S M U S S S S S S S S S S U U U S S S ulfuric A cid 50 M U U U S U U S S M U U S U U M S S S S S U U U M S S ulfuric A cid conc M U U U U U M M...

Page 85: ...d U U U S S U M S U S U U S M M S S U U S U U U M U T richloroethane S U M U U S U U U U U U U U U S U S S T richloroethylene U U U U S U U U U U U U U U S U U S T risodium P hosphate S M S S S S S S...

Page 86: ...ne S S U S S S M U S U U U U U U M U M U U U S U M S U S Z inc C hloride U U S S S S U S S S S S S S S S S S S S S S S U S S S Z inc S ulfate U S S S S S S S S S S S S S S S S S S S S S S S S S C itri...

Page 87: ...7 H Hematocrit Rotor 31 How to install a rotor 40 How to Operate a Rotor 40 How to remove a rotor 41 I Ice Formation 63 Information for the Customer Service 66 Intended Use 5 Items Supplied 32 L List...

Page 88: ...Standards 16 Storage 37 T Technical Data 12 Technical Specifications 11 Transport and Set Up 32 Troubleshooting 62 Troubleshooting by Guide 63 U Unpacking 32...

Page 89: ...isherbrand accuSpin Micro 21R Thermo Electron LED GmbH Zweigniederlassung Osterode Am Kalkberg 37520 Osterode am Harz Germany Distributed by Fisher Scientific Contact us today UK Responsible Person Li...